Using the Astral device
English
33
Exporting device settings onto USB stick
Exporting of device settings can be performed at any time.
To export device settings:
1.
Insert a USB stick into the USB connector at the rear of the device. Refer to Data Management
process (see page 127).
2.
From the Main menu, press Setup
. The Setup menu is displayed.
3.
Select the
Settings
tab from the
USB
sub-menu.
When the USB is ready, the
Export >
button will become active.
4.
Press
Export >
.
Any Astral settings currently on the USB stick will be deleted and replaced with the settings from the current
device.
5.
Press
Confirm
to proceed with the export.
When the export is complete, a status message is displayed.
Importing device settings from the USB stick
Settings can only be imported when the device is in Clinical Mode and not ventilating.
Notes:
•
All settings should be verified to be appropriate for the patient prior to commencing ventilation.
•
Learn circuit must be performed for each enabled program prior to commencing ventilation.
To import device settings:
1.
Insert a USB stick into the USB connector at the rear of the device. Refer to Data Management
process (see page 127).
2.
Enter
Clinical mode
.
3.
From the Main menu, press Setup
. The Setup menu is displayed.
4.
Select the
Settings
tab from the
USB
sub-menu.
The
Import >
button will become active if there are settings from a compatible device available on the USB stick
and the device is not ventilating.
5.
Press
Import >
.
A warning will be displayed to indicate that all device and patient program settings will be replaced.
6.
Press
Confirm
to proceed with the import.
A progress bar will be displayed while the settings are being imported and applied.
When the import is complete, a status message is displayed as well as a reminder to check all settings have
been applied correctly.
Learn Circuit results will be reset to default parameters for each enabled program. Learn Circuit needs to be
performed with the appropriate patient breathing circuit for each enabled program.
All device configuration settings and all patient program settings should be verified to be appropriate for the
patient prior to commencing ventilation.
Summary of Contents for Astral 100
Page 1: ...Clinical guide English ...