Circuit options
English
39
6.
Attach the patient interface (eg, mask) to the Leak valve or the free end of the air tubing as
appropriate and adjust the mask type setting on the Astral device.
Note:
If using the ResMed Leak Valve, select Leak Valve for the Mask type setting.
Connecting a single limb circuit for invasive use
CAUTION
Always set up the ResMed Leak Valve in the breathing circuit with the arrows and the symbol
pointing in the direction of air flow from the Astral device to the patient.
For invasive ventilation, since the patient's upper respiratory system is bypassed by an artificial airway
device (for example endotracheal or tracheostomy tube) humidification of the inspired gas is required to
prevent lung injury.