Ventilation modes
84
Entering the target values manually
The Target Va can also be determined to adopt a Target Patient Rate using a settable Target Va
parameter and patient height. Target Pt Rate should match the patient's normal breathing rate.
To enter the target values manually
1.
From the
Setup
main menu, select
Settings
.
2.
From the
Settings
menu, select the
Main Settings
tab.
3.
From the on-screen selections, press
Ventilation Mode
, then select
iVAPS
.
4.
Select
Skip
.
Summary of Contents for Astral 100