16

Symbols Which Appear On The Device

 Caution; 

 Follow instructions for use; 

 Drip proof; 

 Type CF equipment; 

 Dangerous voltage; 

 Class II equipment; 

 Start/Stop and mask-fit; 

 Manufacturer; 

 

Environmental information 

WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic 
equipment. This device should be disposed of separately, not as unsorted municipal waste. To dispose 
of your device, you should use appropriate collection, reuse and recycling systems available in your 
region. The use of these collection, reuse and recycling systems is designed to reduce pressure on 
natural resources and prevent hazardous substances from damaging the environment. 
If you need information on these disposal systems, please contact your local waste administration. The 
crossed-bin symbol invites you to use these disposal systems. If you require information on collection 
and disposal of your ResMed device please contact your ResMed office, local distributor or go to 
www.resmed.com/environment.

General Warnings and Cautions

!

Warnings

Advice contained in this manual should not supersede instructions given by the 
prescribing physician.

A patient should not connect a device to the data communication port unless instructed 
to do so by their health care provider or physician. Only ResMed products are designed 
to be connected to the data communication port. Connecting other devices could result 
in injury, or damage to the S8 device.

This S8 device should only be used with masks (and connectors

*

) recommended by 

ResMed, or by a physician or respiratory therapist. A mask should not be used unless the 
S8 device is turned on and operating properly. The vent hole or holes associated with the 
mask should never be blocked.

Explanation:

 This S8 device is intended to be used with special masks (or connectors

*

which have vent holes to allow continuous flow of air out of the mask. When the device 
is turned on and functioning properly, new air from the device flushes the exhaled air out 
through the mask vent holes. However, when the device is not operating, insufficient 
fresh air will be provided through the mask, and the exhaled air may be rebreathed. 
Rebreathing of exhaled air for longer than several minutes can, in some circumstances, 
lead to suffocation. This applies to most Positive Airway Pressure devices.

In the event of power failure

 or machine malfunction, remove the mask. 

This S8 device can be set to deliver pressures up to 25 cm H

2

O. In the unlikely event of 

certain fault conditions, pressures up to 40 cm H

2

O are possible.

Follow all precautions when using supplemental oxygen.

Oxygen flow must be turned off when the flow generator is not operating, so that 
unused oxygen does not accumulate within the flow generator enclosure and create a 
risk of fire.

If the oxygen has been left on, turn off the device, then wait 30 minutes before turning on 
the device again.

*

Ports may be incorporated into the mask or in connectors that are near the mask.

During partial (below rated minimum voltage) or total power failure, therapy pressures will not be 
delivered. When power is restored, operation will recommence with no change to settings.

Summary of Contents for Positive AirwAy Pressure Device S8 Auto 25

Page 1: ...S8 Auto 25 Positive Airway pressure device User Guide English ...

Page 2: ...9 AU 724589 AU 764761 EP 0651971 EP 0661071 EP 0858352 EP 0920845 EP 0927538 EP 0934723 JP 3638613 US 5199424 US 5245995 US 5522382 US 5704345 US 6138675 US 6213119 US 6240921 US 6363933 US 6367474 US 6425395 US 6502572 US 6591834 US 6675797 US 6705315 US 6745768 US 6817361 US 6988498 US 7040317 US 7100608 Other patents pending Protected by design registrations AU 302180 AU 302181 AU 302182 AU 302...

Page 3: ...ontrol Panel 4 Using the Menus 4 SmartStart 6 Starting Treatment 7 Stopping Treatment 7 Using the Mask Fitting feature 7 Reminders on the S8 LCD 8 Smart Data 8 Using the Data Card 9 Cleaning and Maintenance 10 Daily 10 Weekly 10 Monthly 10 Replacing the Air Filter 11 Servicing 11 Troubleshooting 12 Technical Specifications 14 Symbols Which Appear On The Device 15 General Warnings and Cautions 15 L...

Page 4: ...lood pressure dehydration cerebrospinal fluid leak recent cranial surgery or trauma Adverse Effects Patients should report unusual chest pain severe headache or increased breathlessness to their prescribing physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during the course of therapy with the device drying ...

Page 5: ... with this S8 device ResScan Data Card The ResScan Data Card may be used with the S8 Auto 25 to help your clinician monitor your treatment or to provide you with updates to your device settings Screen Keypad Air outlet DC AC Air inlet Data Card module Carry handle Mask Type Name Nasal Masks Mirage Vista Nasal Mask Ultra Mirage Nasal Mask Ultra Mirage II Nasal Mask Mirage Activa Nasal Mask Mirage M...

Page 6: ...d and plug are in good condition and the equipment is not damaged Only ResMed air tubing should be used with the device A different type of air tubing may alter the pressure you actually receive reducing the effectiveness of your treatment Blocking the hose and or air inlet of the device while in operation could lead to overheating of the device CAUTION Be careful not to place the device where it ...

Page 7: ...r S8 device includes an LCD screen which displays the menus and treatment screens as well as a keypad for navigating through the menus and delivering treatment The keypad has the following keys LCD screen Keypad Start Stop Key Key Function Start Stop Starts or stops treatment Extended hold for at least three seconds starts the mask fit feature Up Allows you to increase settings options and scroll ...

Page 8: ...iled view The following illustration summarises these views SETTLE 10min menu WELCOME 00006 00012 days USED 00090hrs SW SX459 0200 exit MASK ULTRA change exit SETTINGS enter exit RESULTS enter exit enter exit OPTIONS TUBE LENGTH 2m change exit SMARTSTART OFF exit MASK ULTRA change exit LEAK ALERT ON change exit LANGUAGE ENGLISH change exit SMART DATA enter exit enter exit SERVICING SW SX459 0200 e...

Page 9: ...nician you can adjust the Exhalation rate so that the rate of pressure drop is at a level that is most comfortable for you There are three settings Fast Medium and Slow 1 On the standby SETTLE screen press 2 Press until you see EXHALATION then press 3 Press or until you see the setting that you require 4 Press to save the setting SmartStart If your clinician has enabled SmartStart Stop your device...

Page 10: ...Start Stop simply remove your mask and treatment will end Note SmartStart Stop does not operate if you have a Full Face Mask or Leak Alert is enabled Using the Mask Fitting feature The S8 Auto 25 includes a mask fitting feature to help you fit your mask properly The mask fitting feature delivers air pressure for a three minute period prior to starting treatment for checking and adjusting your mask...

Page 11: ...ll also remove the message REPLACE MASK Reminds you that your mask is due for replacement Press Ok to remove the message from your LCD and replace your mask with a new one CALL PROVIDER Reminds you to contact your clinician for example to discuss your therapy Press Ok to remove the message from your LCD and contact your clinician service provider REPLACE FILTER Reminds you to replace the air filte...

Page 12: ... Return the card in its protective folder to your clinician using a postal envelope Updating Settings on your S8 Auto 25 If your clinician has provided a Data Card with new device settings 1 With the device in standby Ramp mode insert the Data Card into the slot on the Data Card module Updating will start automatically The Card Inserted Please Wait message is displayed on the LCD while updating is...

Page 13: ...hat country Use on an Aircraft Please consult the medical services department of your carrier if you intend to use your S8 device on an aircraft Note You should not use your S8 device while the aircraft is taking off or landing Use with DC Power You must use a ResMed DC DC Converter 24V 50W to connect your S8 Auto 25 to a 12V or 24V DC power source Contact your equipment supplier or ResMed for det...

Page 14: ...e air filter for holes and blockage by dirt or dust WARNING Beware of electric shock Do not immerse the unit or power cord in water Always unplug the unit before cleaning and be sure it is dry before plugging back in The mask system and air tubing are subject to normal wear and tear Inspect them regularly for damage Do not use bleach chlorine alcohol or aromatic based solutions moisturising or ant...

Page 15: ...iable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed Applicable ResMed warranty details are provided with the device at the time of original supply Of course as with all electrical devices if any irregularity becomes apparent you should exercise caution and have the device inspected by an authorised ResMed service centre CAUTION Do not ...

Page 16: ...ated incorrectly causing excessive leak Adjust headgear or re fit cushion Humidifier control dial set too high resulting in accumulation of water in the air tubing Turn humidifier control down and empty the water from the air tubing Breath is not deep enough to trigger SmartStart Stop Take a deep breath in and out through the mask There is excessive leak Adjust position of mask and headgear Air tu...

Page 17: ...Cause Solution Component failure Record error number and contact your ResMed service centre You have experienced excessively high leak levels for more than 20 seconds Check that your air tubing is connected properly Adjust headgear Data Card is not inserted correctly Ensure that the Data Card is inserted with the arrow facing up as far as it can go You may have removed the Data Card before setting...

Page 18: ... level 36 dBA with uncertainty of 2 dBA as measured according to ISO 17510 1 2007 Dimensions L x W x H 112 mm x 164 mm x 145 mm Weight 1 3 kg Power supply Input range 100 240V 50 60Hz 40VA typical power consumption 100VA maximum power consumption Housing construction Flame retardant engineering thermoplastic Operating temperature 5ºC to 35ºC Operating humidity 10 95 non condensing Storage and tran...

Page 19: ...ld only be used with masks and connectors recommended by ResMed or by a physician or respiratory therapist A mask should not be used unless the S8 device is turned on and operating properly The vent hole or holes associated with the mask should never be blocked Explanation This S8 device is intended to be used with special masks or connectors which have vent holes to allow continuous flow of air o...

Page 20: ...in the vicinity of flammable anesthetics Cautions At low pressures the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing Some rebreathing may occur The airflow for breathing produced by this device can be as much as 11ºF 6ºC higher than the temperature of the room Caution should be exercised if the room temperature is warmer than 90ºF 32ºC When AC mains p...

Page 21: ...han those specified in this manual is not recommended They may result in increased emissions or decreased immunity of the S8 device Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices The S8 device is intended for use in an environment in which radiated RF disturbances are controlled The customer or the user of the S8 device can hel...

Page 22: ...ation in a typical commercial or hospital environment Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 10 V m 80 MHz to 2 5 GHz 3 Vrms 10 V m Portable and mobile RF communications equipment should be used no closer to any part of the S8 device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the trans...

Page 23: ...defective product must be made by the initial consumer at the point of purchase This warranty is in lieu of all other express or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for...

Reviews: