Cleaning and maintenance

Daily:

Remove the air tubing by pulling on the finger grips on the cuff. Hang it in a clean, dry place until next 
use.

Weekly: 

1. Remove the air tubing from the S9 device and the mask.
2. Wash the air tubing in warm water using mild detergent.
3. Rinse thoroughly, hang, and allow to dry.
4. Before next use, reconnect the air tubing to the air outlet and mask.

Monthly:

1. Wipe the exterior of the S9 with a damp cloth and mild detergent.
2. Check the air filter for holes and blockage by dirt or dust.

Replacing the air filter:

Replace the air filter every six months (or more often if necessary).

1. Remove the air filter cover from the back of the S9 device.
2. Remove and discard the old air filter.
3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
4. Replace the air filter cover.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your 
equipment supplier or ResMed. Do not attempt to open these devices.

Problem/Possible cause

Solution

No display

Power is not connected.

Ensure the power cable is connected and the power outlet 
(if available) is on.

The DC plug is partially inserted into the 
back of the device.

Fully insert the DC plug.

Insufficient air delivered from the device

Ramp time is in use.

Wait for air pressure to build up or change ramp time.

Air filter is dirty.

Replace air filter. 

Air tubing is not connected properly.

Check air tubing. 

368147 S9 AutoSet S9 Elite User Guide ROW.book  Page 3  Thursday, August 4, 2011  8:57 AM

Summary of Contents for S9 AutoSet

Page 1: ...17 US 7100608 US 7537010 US 7866944 Other patents pending Protected by design registrations AU 329025 AU 330142 AU 331272 AU 332891 AU 333858 CN 200830346524 1 CN 200830346526 0 CN 200830346525 6 CN 201030215797 X EU 1067417 JP 1377401 JP 1377402 JP 1377403 JP 1415251 JP 1415252 NZ 411762 NZ 413090 NZ 413091 US D629891 US D629892 US D635232 US D635244 US D635245 US D635654 US D638536 US D638537 US...

Page 2: ...ogically low blood pressure dehydration cerebrospinal fluid leak recent cranial surgery or trauma Adverse effects Patients should report unusual chest pain severe headache or increased breathlessness to their prescribing physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during the course of therapy with thes...

Page 3: ...u to enter into a menu and confirm your choice Navigating the menus Refer to illustration C In general to navigate the menus 1 Turn until the parameter you require is displayed in blue 2 Press The selection is highlighted in orange 3 Turn until you see the setting that you require 4 Press to confirm your choice The screen returns to blue Getting started 1 Make sure the power is connected 2 Adjust ...

Page 4: ... 2 Remove and discard the old air filter 3 Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover 4 Replace the air filter cover Troubleshooting If there is a problem try the following suggestions If the problem cannot be solved contact your equipment supplier or ResMed Do not attempt to open these devices Problem Possible cause Solution No display Power is not con...

Page 5: ...ble SmartStart Stop SmartStart Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system being used Only use equipment recommended by ResMed The patient is using a nasal pillows mask with a set pressure less than 7 cm H2O Disable SmartStart Stop Pressure rises inappropriately Talking coughing or breathing in an unusual manner Avoid talking with a...

Page 6: ...cked please check your tube Air tubing is blocked Check your air tubing and remove any blockages Displays error message High leak please check system setup and all connections There is excessive leak Adjust position of mask and headgear Air tubing is not connected properly Connect firmly at both ends The following message is displayed on the LCD after you try to update settings or copy data to the...

Page 7: ...ult steady state pressure 30 cm H2O Pressure measurement tolerance 0 5 cm H2O 4 of the measured reading Flow measurement tolerance 6 L min or 10 of reading whichever is greater DECLARED DUAL NUMBER NOISE EMISSION VALUES in accordance with ISO 4871 Sound pressure level 24 dBA as measured according to ISO 17510 1 2002 26 dBA with uncertainty of 2 dBA as measured according to EN ISO 17510 1 2009 Soun...

Page 8: ...lectrical components 1 9 m 19 mm inner diameter Air outlet The 22 mm conical air outlet complies with ISO 5356 1 Electromagnetic compatibility Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 for residential commercial and light industry environments Information regarding the electromagnetic emissions and immunity of these ResMed devices...

Page 9: ...as unsorted municipal waste To dispose of your device you should use appropriate collection reuse and recycling systems available in your region The use of these collection reuse and recycling systems is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment If you need information on these disposal systems please contact your local waste ad...

Page 10: ...se This warranty replaces all other expressed or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale ...

Page 11: ... the device before cleaning and make sure it is dry before plugging it back in Do not use bleach chlorine alcohol or aromatic based solutions moisturising or antibacterial soaps or scented oils to clean the S9 or air tubing These solutions may cause damage and reduce the life of these products Follow all precautions when using supplemental oxygen Explosion hazard do not use in the vicinity of flam...

Page 12: ... device flushes the exhaled air out through the mask vent holes However when the device is not operating insufficient fresh air will be provided through the mask and the exhaled air may be rebreathed Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation This applies to most models of CPAP devices In the event of power failure2 or machine malfuncti...

Page 13: ...ally crack Do not wash the air filter The air filter is not washable or reusable The air filter cover protects the device in the event of accidental liquid spillage Ensure the air filter and air filter cover are fitted at all times At low pressures the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing Some rebreathing may occur The airflow for breathing p...

Reviews: