Technical specifications
The following message is NOT displayed on the LCD after you try to update the settings using
the SD card:
Settings updated successfully, press any key
The settings were not updated.
Contact your clinician/service provider immediately.
Operating pressure range
4 to 20 cm H
2
O
Maximum single fault steady state
pressure
30 cm H
2
O
Pressure measurement tolerance
±0.5 cm H
2
O ± 4% of the measured reading
Flow measurement tolerance
±6 L/min or 10% of reading, whichever is greater
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
Sound power level
34 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
Nominal dimensions (L x W x H)
153 mm x 140 mm x 86 mm
Weight
835 g
90W power supply unit
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 70W (80VA)
Maximum power consumption 110W (120VA)
30W power supply unit
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 20W (30VA)
Maximum power consumption 36W (75VA)