15
Parameter
Description
Learn Circuit Refer to illustration O.
The device is calibrated according to your air tubing system.
1.
Ensure that the therapy is turned off before performing a Learn Circuit.
2.
If in use, turn off the oxygen flow.
3.
Select the mask type.
4.
Set up the air circuit including accessories and patient interface.
Note:
When performing a Learn Circuit for invasive use, do not connect a catheter
mount, tracheostomy tube or HMEF. Refer to Setting up for invasive use (see page 4).
5.
Leave the air circuit unobstructed and open to the air.
6.
Press
to start the Learn Circuit.
7.
Wait for the device to complete its automated tests (<30 sec).
The results are displayed when complete. If the circuit configuration has been
successfully learnt,
displays. If unsuccessful,
displays (refer to Troubleshooting
(see page 19)).
FiO
2
Sensor
Calibration
The device starts the calibration of the FiO
2
monitoring sensor to measure the oxygen
concentration of the breathable air.
1.
Press
to start the FiO
2
sensor calibration.
2.
Wait for the device to complete its calibration.
The results are displayed when complete.
Note:
Turn off the oxygen flow.
Ramp Time
If Max Ramp Time has been set by the clinician, you may select any value up to this time.
Options:
0 min-Max Ramp Time (max 45 minutes, 5 minutes increments)
Setup menu: Configuration Menu
Parameter
Description
Language
Sets the display language.
Options:
Depending on regional configuration
Brightness
Sets the LCD backlight brightness.
Options:
20-100%, 10% increments
Backlight
Enables the LCD and keypad backlight.
If the AUTO setting is selected the backlight turns dark after five minutes without
any actions and turns on again if any button is pressed again or an alarm occurs.
Options:
On, Auto
Time format
Sets the time format.
Options:
24 hrs, 12 hrs