Problem / possible cause
Action
LCD: Non-Vented Mask!
•
Connection of a non-vented
mask.
•
Mask vents may be blocked.
•
ResMed Leak Valve is
missing or vent is blocked.
•
Ensure the mask has vents.
•
Ensure the mask vents are not blocked.
•
Ensure the ResMed Leak Valve is installed and that the vent is not
blocked.
•
Ensure oxygen (if in use) has only been connected at the rear of the
device.
•
If the problem persists, contact your clinician.
LCD: Apnoea!
The device detects an apnoea
that has exceeded the pre-set
alarm level.
•
Breathe normally to disable the alarm.
•
If the problem persists, contact your clinician.
LCD: Internal Battery Low!
The internal battery capacity is
below 30%.
Connect the device to mains power.
LCD: Low SpO
2
!
SpO
2
has dropped below pre-set
alarm level.
•
Check the attachment of the sensor.
•
If the problem persists, contact your clinician.
LCD: SpO
2
finger sensor failure!
The finger sensor is not
connected properly or delivers
faulty values.
Check if the finger sensor is attached properly to the finger and
connection to the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is
disconnected.
Check if the pulse oximeter is connected properly to the device.
LCD: Low FiO
2
Level!
FiO
2
has dropped below the pre-
set alarm level.
•
Perform FiO
2
sensor calibration.
•
If the problem persists, contact your clinician.
LCD: High FiO
2
Level!
FiO
2
has exceeded the pre-set
alarm level.
•
Perform FiO
2
sensor calibration.
•
If the problem persists, contact your clinician.
LCD: FiO
2
sensor disconnected!
The FiO
2
monitoring sensor is not
connected.
Connect the FiO
2
monitoring sensor or replace it with a new one.
LCD: Keypad Failure!
One of the keys was held down
for more than 10 seconds or got
stuck.
Remove any blockages from the keypad.
LCD: Attention High Temp [42, 43, 44, 45]!
The temperature inside the
device is high.
Ensure the ambient temperature conditions are within the specified
operating range.
LCD: Internal Battery Use!
The device is using the internal
battery.
Check if the power cord is properly connected to the device if you
want to run from mains power.
Press the Alarm mute button
to cancel the alarm.