22
Problem / possible cause
Action
LCD: External DC Power Use!
The device is powered by an
external battery.
Check if the AC power cord is properly connected to the device if you
want to run from mains power.
Note:
The alarm will be cleared automatically after one minute.
Other troubleshooting
Problem / possible cause
Solution
No display
Power failure. The device stops delivering
air pressure.
•
Remove the mask or the catheter mount from the
tracheostomy tube until power is restored.
•
Check the power cable connection between the mains
power or external battery and the device.
•
As the internal battery ages, the available capacity
decreases. When the remaining battery capacity is low, it
can affect the battery charge level indication and lead to
power failure without activating Low or Empty battery
alarms. ResMed recommends the battery be tested after
two years to assess the remaining battery life (refer to
Internal battery (see page 2)).
Power not connected or device is not
switched on.
Ensure the power cord is connected and press the switch at
the back of the device once.
Treatment pressure seems low
Ramp time is in use.
Wait for air pressure to build up.
Air filter is dirty.
Replace air filter.
Air tubing is kinked or punctured.
Straighten or replace tubing.
Air tubing is not connected properly.
Connect the air tubing firmly at both ends.
Mask and headgear not positioned
correctly.
Adjust position of mask and headgear.
Plug(s) missing from access port(s) on
mask.
Replace plug(s).
Pressure required for treatment may have
changed.
See your clinician to adjust the pressure.
There is a large impedance (eg,
bacterial/viral filter) in the air circuit.
Perfom the Learn Circuit function.
Humidifier control dial set too high,
resulting in accumulation of water in the
air tubing.
Turn humidifier control down and empty the water from the air
tubing.
Treatment pressure seems high
Pressure required for treatment may have
changed.
Consult your clinician.
There is a change in the impedance in the
circuit configuration.
Perform the Learn Circuit function.
The device does not start when you breathe into the mask
SmartStart/Stop not on.
Consult your clinician.
Breath is not deep enough to trigger
SmartStart.
Take a deep breath in and out through the mask.
There is excessive leak.
Adjust position of mask and headgear.
Plug(s) missing from port(s) on mask.
Replace plug(s).
Summary of Contents for Stellar 100
Page 212: ...2 1 I 2 1 J 1 2 3 4 5 K ...