24
Problem / possible cause
Solution
LCD:
is displayed in the header.
Battery is not charging.
•
Ensure the ambient temperature conditions are within the
specified operating range. If the problem persists within
the specified operating conditions please return the device
for servicing.
•
Power off the device. Power on the device again.
Technical specifications
Operating pressure range
•
IPAP: 2–40 cm H
2
O (in S, ST, T, PAC mode)
•
PS: 0–38 cm H
2
O (in S, ST, T, PAC mode)
•
EPAP: 2–25 cm H
2
O (in S, ST, T, iVAPS, PAC mode)
•
CPAP: 4–20 cm H
2
O (in CPAP mode only)
•
Min PS: 0–20 cm H
2
O (in iVAPS mode)
•
Max PS: 0–30 cm H
2
O; 8–30 cm H
2
O when AutoEPAP is on (in iVAPS mode)
•
Min EPAP/Max EPAP: 2–25 cm H
2
O (in iVAPS mode when AutoEPAP is set to on)
•
Intended tidal volume range: 50–2500 mL
Note:
iVAPS mode is only available in Stellar 150.
Maximum single fault pressure
60 cm H
2
O (in all modes)
Maximum breathing resistance
under single fault
2 cm H
2
O at 30 L/min; 7.2 cm H
2
O at 60 L/min
Maximum flow
> 200 L/min at 20 cm H
2
O
Flow accuracy
± 5 L/min or 20% measured value, whichever is greater
Test condition: T mode, IPAP: 40 cm H
2
O, EPAP: 2 cm H
2
O, Rise Time: MIN, Fall Time: MIN,
Ti: 4.0 sec, Respiratory Rate: 10 bpm, with ResMed calibration cap.
Therapy pressure tolerance
IPAP: ± 0.5 cm H
2
O ± 10% of set pressure (end of inspiration)
EPAP/PEEP: ± 0.5 cm H
2
O ± 4% of set pressure
CPAP: ± 0.5 cm H
2
O ± 10% of set pressure
Test condition: T mode, IPAP: 40 cm H
2
O, EPAP: 2 cm H
2
O, Rise Time: MIN, Fall Time: MIN,
Ti: 4.0 sec, Respiratory Rate: 10 bpm, with ResMed calibration cap.
Sound pressure level
31 dBA with uncertainty of 3 dBA as measured according to ISO 17510-1.
Sound power level
39 dBA with uncertainty of 3 dBA as measured according to ISO 17510-1.
Alarm Volume Range
High, medium and low priority alarm (measured in a 1m radius according to IEC 60601-1-8):
minimum 42 dB
maximum 79 dB
Dimensions (L x W x H)
230 mm x 170 mm x 120 mm
Weight
2.1 kg
LCD screen
Dimension (L x W x H): 76.9 mm x 63.9 mm x 3.15 mm
Resolution: 320 x 240 pixels
Air outlet
22 mm taper, compatible with ISO 5356-1 Anaesthetic & Respiratory Equipment - Conical
Connectors
Pressure measurement
Internally mounted pressure transducer