14
Setup menu
Setup menu: Clinical Settings (Mask Type)
1.
Press
to display the
Clinical Settings
screen.
2.
Use the push dial
to scroll through the menu and change the
Mask Type
in the
Advanced Settings
screen.
Mask type options: Nasal, Ultra, Pillow, Full Face, Trach, Pediatric.
Note:
When Mask Type is set to Trach or Full Face the Non-Vented Mask alarm automatically turns on.
For a full list of compatible masks for this device, see the Mask/Device Compatibility List on
www.resmed.com/downloads/devices. If you do not have internet access, please contact your ResMed
representative.
Setup menu: Alarm Settings (Alarm Volume)
1.
Press
to display the
Alarm Settings
screen.
2.
Use the push dial
to scroll through the menu and change the
Alarm Volume
to low, medium or
high.
Setup menu: Options
1.
Press
to display the
Options
screen.
2.
Use the push dial
to scroll through the menu and change parameters (see descriptions in the table
below)
.