Troubleshooting guide 

Airlink 2 not found  

CAUSE 

POSSIBLE REMEDY 

No Airlink 2 inserted in the PC’s USB2 port. 

Insert Airlink 2 and launch the 

fi

tting software if not 

already launched. 

Airlink 1 dongle inserted and is being used instead. 

Remove the Airlink 1 dongle from the USB port. 

 

No Connection 

CAUSE 

POSSIBLE REMEDY 

The USB port(s) in your PC have no power. 

Not an Airlink 2 issue. PC service is required. 

No Airlink 2 attached to the PC's USB port. 

Insert Airlink 2 and launch the 

fi

tting software if not 

already launched. 

Obstacles between the Airlink 2 and the hearing 
instruments block the wireless signal. 

Place the Airlink 2 in line of sight of the hearing 
instruments. 

The Airlink 2 is placed too far away from the hearing 
instruments. 

Move the Airlink 2 closer to the hearing 
instruments. 

Low battery power

.

 

Always use fresh zinc-air batteries, which have not 
expired. Remove the sticker from the battery and 
let the battery breathe for 2 minutes before 
inserting and starting up the fitting. 

 

Unstable/Lost Connection 

CAUSE 

POSSIBLE REMEDY 

Obstacles between the Airlink 2 and the hearing 
instruments block the wireless signal

.

 

Place the Airlink 2 in line of sight of the hearing 
instruments. 

The Airlink 2 is placed too far away from the hearing 
instruments. 

Move the Airlink 2 closer to the hearing 
instruments. 

A 2.4 GHz cordless telephone, a WLAN access point 
or another wireless device is powered or in use near 
to the 

fi

tting station doing wireless 

fi

tting with 

Minimize number of interfering wireless sources or 
move them away from the fitting area. 

Summary of Contents for Airlink 2

Page 1: ...0 USER GUIDE Airlink 2...

Page 2: ...reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment gene rates uses and can radiate radio frequency energy and if not installed...

Page 3: ...orme NMB 003 du Canada Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may caus...

Page 4: ...less fitting 5 Compatibility 7 Light indicators LED 8 Important information 9 Technical specifications 9 Warranty 10 Trouble shooting guide 11 Description 1 Left side light indicator 2 Right side ligh...

Page 5: ...t you have the appropriate fitting software installed The Airlink 2 driver is installed during fitting software installation Airlink 2 is supported by ReSound Aventa 3 6 or later Airlink 2 is supporte...

Page 6: ...ting wireless ReSound Beltone and Interton products using ReSound Aventa Beltone Solus Pro and Interton Appraise Tips for optimal wireless fitting Let the battery breathe oxygenize before inserting Wh...

Page 7: ...be impacted by interference from other wireless sources in the environment Interference may cause data transfer issues slow connectivity or loss of connections to the hearing instruments entirely Inte...

Page 8: ...3 6 or later Airlink 2 works with all ReSound s wireless WL hearing instrument models Airlink 2 is compatible with Beltone Solus Pro 1 6 or later Airlink works with all Beltone s wireless WL hearing i...

Page 9: ...300ms On 300ms Off If data transfer to Right Blinking green 300 ms On 300 ms Off Airlink connected to PC FSW launched HI data transfer HI upgrade in progress If connection to Left HI is lost Blinking...

Page 10: ...e move the Airlink 2 away from the source of interference Airlink 2 must not be used for any other purpose than for hearing instrument fitting General precautions Your hearing instrument and this devi...

Page 11: ...er for a period of 12 months from the date of original purchase Please notice that extended warranties may apply in your country Please contact your local hearing instrument manufacture Be aware of in...

Page 12: ...Airlink 2 is placed too far away from the hearing instruments Move the Airlink 2 closer to the hearing instruments Low battery power Always use fresh zinc air batteries which have not expired Remove...

Page 13: ...In order to minimize environmental pollution dispose the product according to local standards and national regulations Any issues relating to the EU Medical Device Directive 93 42 EEC and R TTE Direc...

Page 14: ...13 Back Page GN Hearing A S Lautrupbjerg 7 DK 2750 Ballerup Denmark Tel 45 45 75 11 11 Fax 45 45 75 11 19 18956400 GB 13 06 Rev A...

Reviews: