56
57
RIE - HP receiver
Models: RE961-DRWC, RE761-DRWC, RE561-DRWC
Frequency (Hz)
Gain (dB)
90
80
70
60
50
40
30
Reference test gain
60 dB SPL input
Full-on gain
50 dB SPL input
2cc Coupler
Frequency (Hz)
Output (dB SPL
)
140
130
120
110
100
90
80
2cc Coupler
Data in accordance with ANSI S3.22-2014, IEC 60118-0:2015
FULL-ON AND REFERENCE TEST GAIN
MAXIMUM OUTPUT (OSPL 90)
RIE - UP receiver
Models: RE961-DRWC, RE761-DRWC, RE561-DRWC
Frequency (Hz)
Gain (dB)
90
80
70
60
50
40
30
Reference test gain
60 dB SPL input
Full-on gain
50 dB SPL input
2cc Coupler
Frequency (Hz)
Output (dB SPL
)
140
130
120
110
100
90
80
2cc Coupler
Data in accordance with ANSI S3.22-2014, IEC 60118-0:2015
FULL-ON AND REFERENCE TEST GAIN
MAXIMUM OUTPUT (OSPL 90)
Reference test gain (60 dB SPL input)
HFA
40
dB
Full-on gain (50 dB SPL input)
Max.
HFA
65
57
dB
Maximum output (90 dB SPL input)
Max.
HFA
120
117
dB SPL
Total harmonic distortion
500 Hz
800 Hz
1600 Hz
0.3
0.7
0.5
%
Equivalent input noise, w/o Noise reduction
22
dB SPL
1/3 Octave Equivalent input noise,
w/o Noise reduction
1600 Hz
10
dB SPL
Frequency range IEC 60118-0: 2015
100-6750 Hz
Expected operating time*
30
Hours
Reference test gain (60 dB SPL input)
HFA
47
dB
Full-on gain (50 dB SPL input)
Max.
HFA
75
65
dB
Maximum output (90 dB SPL input)
Max.
HFA
128
124
dB SPL
Total harmonic distortion
500 Hz
800 Hz
1600 Hz
1.0
1.6
0.1
%
Equivalent input noise, w/o Noise reduction
21
dB SPL
1/3 Octave Equivalent input noise,
w/o Noise reduction
1600 Hz
9
dB SPL
Frequency range IEC 60118-0: 2015
100-4920 Hz
Expected operating time*
30
Hours
*Expected operating time of the rechargeable battery depends on active features, the
use of wireless accessories, hearing loss, battery age and sound environment.
*Expected operating time of the rechargeable battery depends on active features, the
use of wireless accessories, hearing loss, battery age and sound environment.