Troubleshooting

Problems

Solutions

Compressor overheats.

Turn the unit OFF for a minimum of 40
minutes to allow the compressor to cool.

The compressor does not operate.

Check all power sources to the compressor.
Push the ON/OFF button again.

If you checked all the items above and 
the unit still does not work, allow the
compressor to cool down for a minimum
of 40 minutes before turning the
compressor on again.

The device is noisy.

Should the device still not operate properly after checking the unit as indicated
above, contact Respironics Customer Service at 1-800-345-6443 or 1-724-387-4000.

Check that the filter cover is correctly
positioned.

Medication does not nebulize properly.

Check for proper connection of the 
tubing adapter.

Make sure the nebulizer is not empty.

Clean the nebulizer as described in 
the manual.

Replace the nebulizer. 

Check the intake filter and replace it if it 
is dirty.

Specifications:

Power Supply:

8.5VDC, 1.4A Max

Weight:

1.0 lbs. (with battery)
0.66 lbs. (without battery)

Size: 

3" x 2-5/8" x 1-7/8" (without battery)
3" x 2-5/8" x 2-1/2" (with battery)

Maximum pressure:

17 PSI, 120 KPa

Maximum flow rate:

5 LPM

Mode of operation

20 minutes ON/40 minutes OFF 

Class II Device (Double Safety Insulation).

Type B Device (device with specific protection against electrical hazards).

Certification:

Electric Safety Standards EN60601-1

Respiratory therapy equipment. Nebulizing systems and their components EN13544-1

Electromagnetic Compatibility according to EN60601-1-2

Storage Conditions:

Temperature:

MIN -4°F (-20°C) – MAX 140°F (60°C)

Humidity:

MIN 10%RH – MAX 95%RH

Operating Conditions:

Temperature:

MIN 41°F (5°C) – MAX 104°F (40°C) (with tubing)
MIN 41°F (5°C) – MAX 77°F (25°C) (handheld)

Humidity:

MIN 10%RH – MAX 95%RH

Replacement Parts and Accessories:

Filter 1041433

AC power/battery charger 

1041517

DC  Car Adapter (12V input/8.5V output)

1041515

Rechargeable Lithium Polymer Battery 

RDD499

Pediatric Face Mask 

1048154

Adult Face Mask 

1048285

Micro Plus Nebulizer 

1041431

Symbol Glossary:

Attention: Read accompanying documents

ON/OFF (power)

Alternating Current

Serial Number

~

SN

Optional Rechargeable Battery:

Operating time with the battery 

Up to 60 minutes

Number of treatments per charge

Up to 4 treatments when nebulizing 3 ml of medication

Normal time to charge

Approximately 3.0 hours

Charge cycle

Approximately 400 cycles

Electromagnetic Information:

Portable and RF communication devices such as cellular phones, pagers, etc. can interrupt operation of electrical 
medical equipment. For this reason, your compressor must be placed far enough away from these devices to prevent
interference. This device meets IEC60601-1-2 standard for electromagnetic compatability (EMC).

EMC data sheets (1047862) are available upon request from Respironics, Inc. customer service at 
1-800-345-6643 or 1-724-387-4000 or electronically at www.microelite.respironics.com

1041430_MicroElite_IFU_FNL.qx  12/4/07  3:55 PM  Page 6

Summary of Contents for MicroElite

Page 1: ...nnels de la santé dûment agréés ou sur leur ordonnance Contenu 1 Compresseur MicroElite avec mode d emploi 1 Nébuliseur avec tubulure PRECAUCIÓN Las leyes federales EE UU limitan la venta de este dispositivo a un médico o profesional sanitario con licencia o bajo sus órdenes Contiene 1 Compresor MicroElite con instrucciones 1 Nebulizador con tubos Instructions for Use Mode d emploi Instrucciones p...

Page 2: ...ND CAUTIONS before using your MicroElite WARNINGS AND CAUTIONS DANGER To reduce the risk of electrocution Always unplug this compressor immediately after using Do not place or store product where it can fall or be pulled into a tub or sink Do not place in or drop into water or other liquid Do not reach for product that has fallen into water Unplug immediately WARNING To reduce the risk of burns el...

Page 3: ...k or DC car adapter carefully 5 Do not operate the unit without the filter or with a clogged dirty filter Operating the unit with a dirty filter can impair proper operation and cause damage that is not covered by the warranty 6 Wash your hands before beginning your treatment 7 Assemble your nebulizer and add your prescribed medicine following your prescription and the separate nebulizer instructio...

Page 4: ...tory therapy equipment Nebulizing systems and their components EN13544 1 Electromagnetic Compatibility according to EN60601 1 2 Storage Conditions Temperature MIN 4 F 20 C MAX 140 F 60 C Humidity MIN 10 RH MAX 95 RH Operating Conditions Temperature MIN 41 F 5 C MAX 104 F 40 C with tubing MIN 41 F 5 C MAX 77 F 25 C handheld Humidity MIN 10 RH MAX 95 RH Replacement Parts and Accessories Filter 10414...

Page 5: ...er votre médecin traitant ou appeler le service clientèle de Respironics au 1 800 345 6443 ou au 1 724 387 4000 Lisez ces AVERTISSEMENTS ET MISES EN GARDE avant d utiliser votre MicroElite AVERTISSEMENTS ET MISES EN GARDE DANGER Réduction du risque d électrocution Débranchez toujours ce compresseur immédiatement après usage Ne placez pas ou ne rangez pas ce produit dans un endroit où il pourrait t...

Page 6: ...nt alternatif branchez ce dernier dans la prise à l arrière du compresseur puis branchez l autre extrémité dans la prise de courant électrique 3 Pour utiliser le compresseur avec l adaptateur de courant alternatif branchez ce dernier dans la prise à l arrière du compresseur puis branchez l autre extrémité dans la prise de courant électrique 4 L appareil peut également être utilisé avec le bloc pil...

Page 7: ...ture 41 F 5 C MIN 104 F 40 C MAX avec la tubulure 41 F 5 C MIN 77 F 25 C MAX unité portative Humidité 10 HR MIN 95 HR MAX Pièces de rechange et accessoires Filtre 1041433 Alimentation c a chargeur de pile 1041517 Adaptateur c c pour véhicule entrée 12 V sortie 8 5 V 1041515 Pile au lithium polymère rechargeable RDD499 Masque facial pour enfants 1048154 Masque facial pour adultes 1048285 Nébuliseur...

Page 8: ...IAS Y PRECAUCIONES antes de usar su MicroElite ADVERTENCIAS Y PRECAUCIONES PELIGRO Para reducir el riesgo de electrocución Siempre desenchufe este compresor inmediatamente después del uso No coloque ni guarde el producto donde pueda caer en una bañera o lavabo No coloque ni deje caer en el agua u otros líquidos No trate de tocar el producto que se haya caído dentro del agua Desenchufe inmediatamen...

Page 9: ...Esta unidad ha sido diseñada para el uso intermitente 20 minutes ENCENDI DA 40 minutos APAGADA Para evitar el riesgo de posible choque eléctrico el nebulizador debe funcionar usando la fuente eléctrica especificada No retire la cubierta del compresor ni desarme la unidad excepto para cambiar el filtro No hay piezas que puedan recibir servicio Nunca use la unidad mientras esté dormido o somnoliento...

Page 10: ...pia respiratoria Sistemas nebulizantes y sus componentes EN13544 1 Compatibilidad electromagnética de acuerdo con EN60601 1 2 Condiciones de almacenamiento Temperatura MÍN 4 F 20 C MÁX 140 F 60 C Humedad HR MÍN 10 HR MÁX 95 Condiciones de funcionamiento Temperatura MÍN 41 F 5 ºC MÁX 104 F 40 ºC con tubos MÍN 41 F 5 ºC MÁX 77 F 25 ºC manual Humedad MÍN 10 de HR MÁX 95 de HR Piezas de repuesto y acc...

Page 11: ...des techniciens non agréés et elle ne couvre pas non plus les frais de main d œuvre Tous les objets renvoyés doivent être conditionnés correctement et expédiés en port payé par le distributeur responsable de la maintenance de l appareil Respironics Inc ne sera pas responsable vis à vis de l acheteur ou de tiers pour cause de perte d utilisation de l équipement ou pour tout dommage indirect accesso...

Reviews: