The Mixer Mill MM 500 control
18
Electrical specification
200
– 240 VAC 50/60 Hz
Apparent power
750 VA
Protection rating
IP 30
Electromagnetic compatibility
IEC 31326-1:2012
EMC class according to DIN EN
55011:2009
EN61000-4-2
EN61000-4-3
EN61000-4-5
EN61000-4-6
EN61000-4-11
Hydraulic specifications
Permissible operating pressure
0 to 5 bar
Typical operating pressure of the through-flow
cooling units
1 to 2 bar
Permissible operating pressure for liquid
nitrogen
1.2 to 1.4 bar
Internal channel volume
approx. 0.32 L
Number of hydraulic connections
4
Number of inlets
2
Number of returns
2
Connection thread at the device inlet
Internal thread G ¼"
Connection to the sample cooling unit (chiller,
cryostat)
via hose set (Art. No. 02.707.0188)
Connection thread hose set
External thread G1/4"
External thread G 3/8"
Is the operation as an open system possible?
Yes
Is the operation as a closed system possible?
Yes
Materials in contact with liquid (material
incompatibilities)
Stainless steel 1.4305
Stainless steel 1.4301
Steel galvanized
Nickel-plated brass
Thermal specifications
Can the grinding stations be cooled
individually?
yes
Are comparative measurements
cooled/uncooled possible?
yes
Number of grinding stations that can be cooled
2
Number of grinding stations
2
Largest temperature-controlled grinding jar
volume
2x 125 ml
Smallest temperature-controlled grinding jar
volume
1x 50 ml
Position of the temperature measurement
Thermal plate underneath the grinding jar
Number of temperature sensors
2
Max. permissible deviation right/left
Configurable: Off / 5 / 10 / 15 / 25 °C