Packaging, Transport and Installation
40
PLEASE NOTE:
The use of an air dehumidifier is recommended.
Fig. 25:
Requirements for the installation site
−
Maximum relative humidity < 80 % (at ambient temperatures ≤ 31 °C)
For ambient temperatures UT between 31 °C and 40 °C, the maximum humidity value LF
decreases linearly according to LF = -(UT - 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N9.0015
Humidity
High relative humidity
−
Electronic and mechanical components may be damaged.
−
The performance data alter to an unknown extent.
•
The relative humidity in the vicinity of the device should be kept as
low as possible.
−
Installation height: max. 2
000 m above sea level
The MM 500 control must be set up on a solid and firm surface, because otherwise the
vibrations of the device will be transmitted to the immediate vicinity during the grinding process.