DIRECTIONS FOR USE 

 
WASHING & DISINFECTION OF THE WASTE WATER TANK & PIPING S 

 

Filling of the disinfectant tank : 

 

  

   

 
Remove the waterproof white cap from the fill opening, then thanks to a shower head or the optional set of filling, fill the 
disinfectant tank with water until the LED indicator ” FULL” ignites. 
 

  

 

 

 

Cleaning and disinfection : 

 

  

 

 

Bring the Litoo Shower in bed system to a shower tray then open the draining valve of the waste water tank. 
 
Insert the suction pipe up to the bottom of the disinfectant tank then press on the On/off switch. 
 
During around thirty seconds, the disinfectant solution circulates in all the pipes of the circuit of waste water. 
A 360° rotary jet allows sprinkling the disinfectant solution on the walls of the waste water tank in order to clean and to 
disinfect. 
 
When the tank is empty, the red LED indicator ignites. 
 
Nevertheless, wait as long as some liquid circulates in the piping. 
 
Then switch off the pump. 
 
 

13/16 

Take the TecCare Control disinfectant and fill the measure until the line. Then tip out the product into the  disinfectant tank. 

Summary of Contents for 3018.00

Page 1: ...A4AN3018 doc 27 02 2019 EN Instruction manual SHOWER IN BED SYSTEM LITOO 3018 00 ...

Page 2: ...S 8 RECHARGE OF THE BATTERY 8 FILLING OF THE FRESH WATER TANK 9 FITTING SHOWER SHEET 9 SHOWERING 10 DIGITAL SHOWER THERMOMETER 10 DRAINING MATTRESS 10 DRAINING TANKS TOWARDS SHOWER TRAY 11 DRAINING TANKS TOWARDS TOILET 11 CLEANING THE WASTE WATER CIRCUIT FILTER 12 WASHING DISINFECTION OF THE WASTE WATER TANK PIPING 13 MAINTENANCE 14 PERIODIC INSPECTION 15 MAINTENANCE LOGBOOK 16 1 16 ...

Page 3: ...handle 8 Shower handset with digital thermometer 9 Suction pipe 10 Waste water tank draining valve 11 Fresh water tank draining valve 12 Shower flexible 13 Tilting foot control 14 Ø150mm wheels 15 Ø65mm directional wheels 16 Mesh filter for suction piping 17 Drain to toilet Outlet 18 Drain to toilet Control 2 16 1 7 2 3 4 5 6 14 15 12 8 9 10 11 13 16 17 18 ...

Page 4: ...Water levels Indicators in the fresh water tank Water levels Indicators in the disinfectant tank Shower pump On Off switch Led indicator Draining pump On Off indicator double functions suction of the waste water from mattress and disinfection of the waste water tank and piping ...

Page 5: ...DIMENSIONS 4 16 ...

Page 6: ...n the cleaning shower head with visual and sound alarm in case of overtaking of temperature PERFORMS Total weight 18kg battery included Fresh water tank capacity 22 litres Water volume for shower 17 litres Water volume of reserve 6 5 litres Shower running time 4 min 35s Waste water tank capacity 22 litres Number of complete cycles including shower suction disinfection for a fully charged battery 5...

Page 7: ...vided with a third tank of a 5 litres capacity intended for the disinfection of the complete circuit of waste water The filling of this disinfectant tank is manually made through the filling opening yellow sticker A shower pump allows to suck up of the fresh water and a shower head allows giving a shower to the patient A drain pump allows to suck up of the waste water of the mattress through a fle...

Page 8: ...e the Litoo Shower in Bed system before transferring the individual to be showered Always check the water temperature before applying the handheld shower to the individual Thoroughly clean the Litoo Shower in Bed system after each use Thoroughly disinfect the shower sheet before each use Always check that the drainage valve is closed before filling the fresh water tank Always check that the waste ...

Page 9: ...f the charger to a main power socket Charge cycle and LED indications LED MODE YELLOW Battery not connected YELLOW Battery initialisation and analysis ORANGE Fast charge GREEN with intermittent YELLOW flash Top Off Charge GREEN Trickle Charge Alternating ORANGE GREEN Error How to connect exchangeable DC output plugs 8 16 Fill the disinfectant tank with water TecCare Control Concentrate disinfectan...

Page 10: ...K Remove the waterproof white cap from the fill opening then thanks to a shower head or the optional set of filling fill the fresh water tank in the desired temperature until the Led indicator FULL ignites FITTING SHOWER SHEET Put the shower sheet over the mattress underneath the individual Fix the shower sheet to the bed using the 4 spring clamps supplied 9 16 ...

Page 11: ...the mattress and if possible for a place where forms a pocket of water Switch on the drain pump by pushing on the On Off button A LED indicator ignites Waste water is pumped from the shower sheet to the waste water tank NOTE Before switching off the pump check the flexible hose does not contain residual water 10 16 Usage When the shower pump is activated the display shows the water temperature If ...

Page 12: ...WARDS TOILET Proceed as follows 1 Position the drain flexible hose of evacuation from the outlet n 17 in the lavatory pan 2 Link the suction flexible hose on the draining hose of waste water tank 3 Position the button to the mode draining towards toilet 4 Open the red valve of the draining pipe The picture above shows the red valve in the closed position 5 Switch on the suction pump until the comp...

Page 13: ...FOR USE CLEANING THE WASTE WATER CIRCUIT FILTER To remove the filter simply unscrew the transparent cover counterclockwise Then rinse the different parts of the filter and more particularly the filter screen 12 16 ...

Page 14: ...tank Insert the suction pipe up to the bottom of the disinfectant tank then press on the On off switch During around thirty seconds the disinfectant solution circulates in all the pipes of the circuit of waste water A 360 rotary jet allows sprinkling the disinfectant solution on the walls of the waste water tank in order to clean and to disinfect When the tank is empty the red LED indicator ignite...

Page 15: ...s and liquid soaps The shower sheet must be washed at high temperature by the laundry department of the establishment between two patients to avoid any risk of infection DISINFECTION OF THE WASTE WATER CIRCUIT To see procedure NOTE The flexible hoses are autoclaves easily removed for replacement 14 16 For the disinfection of the covers control panel Reval recommends using TecCare Control disinfect...

Page 16: ...ible for the servicing maintenance of the trolley 4 The date of corrective actions should also be noted in the logbook 5 If the periodic inspection reveals any defect wear or other damage that compromises the safety of the trolley the owner should immediately be alerted In the event of immediate danger the trolley should immediately be removed from service and not used until repaired 6 Defects and...

Page 17: ...MAINTENANCE LOGBOOK Agent name and signature Corrections Observations Operation Safety parts Control date 16 16 ...

Page 18: ...Soulle Tel 33 05 46 42 04 16 Fax 33 05 46 43 06 38 www groupe reval com UNITED KINGDOM Reval Continuing Care Limited Unit 17 Brynmenyn Busines Centre St Theodore s Way Brynmenyn CF32 9TZ BRIDGEND WALES Tel 01656 721 001 Fax 01656 720 006 www revalcc co uk ...