background image

                                                                                          

 

 
 

EN   Instruction manual 
 

 

 

HYDRAULIC STRETCHER LIFT 

 

TOPAZ 

 
5472.00 (hydraulic version) 
5479.00 (electric version) 
 

 

 

 

Summary of Contents for TOPAZ 5472.00

Page 1: ...EN Instruction manual HYDRAULIC STRETCHER LIFT TOPAZ 5472 00 hydraulic version 5479 00 electric version...

Page 2: ...CIFICATIONS 3 OPERATION 3 RANGE OF APPLICATION 4 TRANSFERTS FROM THE BED 5 TO THE BATHTUB 5 PERIODIC INSPECTION 6 MAINTENANCE 7 RECHARGING OF BATTERIES 7 SAFETY SYSTEMS 7 CHANGE OF FUSES 7 MODIFICATIO...

Page 3: ...GENERAL DRAWINGS DIMENSIONS 2 9 60 5cm 85cm 49cm 70cm 84cm 50cm 116cm 5472 00 84cm 133cm 60cm 70cm 50cm 60cm 9 5cm 69cm 5479 00 66cm 40cm 66cm 140cm 42cm 46cm 21cm 69cm 50cm 50cm 12 5cm 9 5cm...

Page 4: ...electric version 60 cm Maximum weight of the patient 150 kg Weight of the stretcher hydraulic verion 58 kg Weight of the stretcher electric version 63 kg CONFORMITY E C marking Equipment in conformit...

Page 5: ...ty of 150kg Transport of patient may be done only if the ground is firm flat and have an inclination of less than 5 degrees The Topaz does not be used during recharging of batteries Only use loader av...

Page 6: ...the patient on the stretcher Adjust the patient s back if necessary raising the jointed extremity of the stretcher Fold the 2 safety bars back On the return to the bed make the reverse order TO THE BA...

Page 7: ...son s responsible for the servicing maintenance of the stretcher lift The date when corrective actions are taken in response to noted observations should also be noted in the logbook 4 A record of the...

Page 8: ...r lift is equipped with two safety systems An emergency stop button avoiding the chair to go up and down An emergency lowering button that s an independent system hand operated located on the frame co...

Page 9: ...he seat Protect the mast of the lift with a plastic film One of the two people has to hold the stretcher during the work Unscrew the cover of the stretcher protection cover of shoulder with a flat scr...

Page 10: ...MAINTENANCE FORM LOGBOOK Agent name and signature Corrections Observations Tests 150kg Stretcher Control date 9 9...

Page 11: ...ulle Tel 33 05 46 42 04 16 Fax 33 05 46 43 06 38 www groupe reval com UNITED KINGDOM Reval Continuing Care Limited Unit 17 Brynmenyn Busines Centre St Theodore s Way Brynmenyn CF32 9TZ BRIDGEND WALES...

Reviews: