Operation and Maintenance Manual – Steam sterilizers mod. Full Clave
Rev. 01 of 01-12-2006 - C:\Documenti\Manuali di istruzione\FULL CLAVE GB-R01.doc
Page 24 of 34
10.4 Extraordinary maintenance to be carried out each month
Press the red key on the tap (SC3) and release the pipe that connects the drain tank (14) to the machine.
(Similarly to all maintenance operations, remember that this must be carried out when the
machine is cold).
Completely remove the drain tank (14) from the machine, unscrew its plug and wash it thoroughly
under running water. Retighten the plug and carefully fit the tank back in its original place. Connect the
drain pipe that projects from the machine back into the tank tap (SC3) and make sure that its red key
trips.
Clean all the exterior parts of the autoclave with a slightly damp sponge, including the door.
Lubricate the pins and closing mechanisms.
Check to make sure that the hinge and door closing pin operate correctly and lubricate them with
Vaseline oil.