Operation and Maintenance Manual – Steam sterilizer mod.  Sterilix Baby 

Rev. 00 del  03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI 
ISTRUZIONE\Baby GB- R00.doc 

 

Pag. 10 of 26 

 

 

5.2

 

Installation and preliminary inspections 

 

 

 

The installer is obliged to make sure that the place in which the sterilizer is installed 
is fit for that purpose (CEI 62/4).  

 

 

The machine is sent packed on a wooden pallet. Make sure that the packing has not been damaged and 
that  everything  is  in  a  perfect  condition.  Handle  the  box  by  means  of  its  handles  and  place  it  on  a 
surface in the immediate vicinity of the place in which the machine is to be positioned (the weight of 
the model is given in chap. 2, in the technical specifications). 
Remove the machine from its wrapping by pulling it out of the box vertically, using 2 hooks (available 
on the market) inserted into the two holes at the sides of the casing. Remove the polystyrene filler balls 
from around the machine. Make sure that none of the parts has been damaged and that the machine is 
complete with all the accessories indicated in the consignment note. Immediately contact our reseller if 
discrepancies are discovered. 
All the packing materials must be kept for guarantee purposes. Machines returned without their original 
packing will not be accepted. 
Install the autoclave in a well ventilated place, well away from sinks, heat sources, grinders and from 
all other machines that create vibrations and/or dust. 

 

Place the autoclave delicately on to a perfectly flat surface. Take care to leave a space of about 5-10 cm 
all around the machine to allow for aeration. 
Autoclave  must  be  installed  in  a  well  lighted  place  equipped  with  an  electrical  system  that  complies 
with the current standards in force. 
Remember that cannot be built into other furniture as it is filled with water from the top. 

 

Fit  the  electric  plug  into  a  socket 

WITH  A  CAPACITY  OF  AT  LEAST  10  Ampere  AND 

EQUIPPED  WITH  AN  EARTH  CONNECTION

.  Make  sure  that  the  mains  voltage  rating 

corresponds to the value indicated on the label at the rear of the machine (230 Volts). 

 

M

ust only be filled with 

DISTILLED WATER

 to ensure that the electrical and air-operated devices 

operate efficiently and are long lasting. 

 

Carry  out  a  cycle  of  dry-run  tests  using  one  indicator  (type  TERMALOG  or  VAPOR  LINE 
EXTENDER). 
Keep the receipt which should be attached to the installation form and sent off along with the warranty. 
 
 

 
 

 

Never connect the plug of the machine to a reduction. 

 

Do not install the  machine in places at altitudes of  more than 1200 m. 
above sea level. 

 

If this is the case, contact the manufacturer for a special version of the 
machine. 

 

Use a bubble level to make sure that the surface on which the appliance is 
to be positioned is perfectly horizontal. 

 

Fill out and send off the warranty form.

 

 
 
 

Summary of Contents for M 9010/56

Page 1: ...Cod 4073 Operation and maintenance manual STEAM AUTOCLAVE MODEL STERILX BABY GB...

Page 2: ...d out by the technician when he installs the machine This form will be considered the reference certificate that validates the warranty It must be sent back to us along with the printer coupon and war...

Page 3: ...uge warning lights and by the position of the controls SYMBOLS AFFIXED TO THE MACHINE WARNING IDENTIFIES A HEAT SOURCE A PART COULD BE DANGEROUSLY HOT WARNING IDENTIFIES A GENERIC DANGER THERE COULD B...

Page 4: ...Five minute final drying phase Switch to select the 121 or 134 cycle Fully visible water filling from the top Built in drying cycle with evaporation system Continuous venting of the air in the chambe...

Page 5: ...ged by a skilled electrician The machines have been designed to operate up to an altitude of 1200 mm above sea level Contact the manufacturer if the machine must operate at a higher altitude as specia...

Page 6: ...ne to prevent accidental contact with hot parts Never ever operate the appliance unless the power cable is in a perfect condition and the structure undamaged It is advisable to keep the door half open...

Page 7: ...rticular preparations are required to operate the machines Just fit the plug into a 230 V power socket The power draw of the machines is 1500 Watt The control panel is elegantly designed easy to read...

Page 8: ...MANUAL 1 10 LITER CAN 1 250 CC DISPENSER DESCRIPTION OF THE REAR PART 12 Power cable with Shuko France plug 13 Fuse T10A 5x20 14 Regolation thermostat 15 Safety valve discharge 16 Used water drain tu...

Page 9: ...iately notify our dealer if discrepancies are discovered Keep the packing materials for as long as the machine remains in use not just for the warranty period This is because the autoclave may need to...

Page 10: ...ll the autoclave in a well ventilated place well away from sinks heat sources grinders and from all other machines that create vibrations and or dust Place the autoclave delicately on to a perfectly f...

Page 11: ...er having made sure that the pressure gauge 6 is on the setting k The machine switches off automatically at the end of the cycle and issues an acoustic signal l If the instruments are still wet after...

Page 12: ...ruments must be sterilized it is advisable to cover the trays with a paper or fabric napkin to prevent direct contact with the tray surface do not place the instruments straight on to the trays 8 Scis...

Page 13: ...1 2 2 1 2 2 2 1 2 2 1 1 1 2 Drying minutes 5 5 5 5 5 Sterilizing minutes with machine hot 23 33 18 23 23 Sterilizing minutes with machine cold 25 35 20 25 25 Max load Kg 1 1 1 1 0 5 Materials metallic...

Page 14: ...mply with the sterilizing parameters indicated by the current standards in force If the autoclave is used to sterilize instruments contaminated by CREUTZFELDT JAKOB s disease use the 134 cycle and ste...

Page 15: ...has tripped Press the reset alongside the control panel Switch 7 faulty Replace the switch Insufficient pressure for the selected cycle with indicator lights 2 and 3 permanently on The bulb thermostat...

Page 16: ...leave the door ajar and store it at a room temperature of 8 C Use a damp cloth to clean the casing of the machine Never ever use inflammable liquids for any reason whatsoever 10 2 Daily routine maint...

Page 17: ...autoclave to operate in a regular way the manufacturer advises users to carry out a functional test to ensure that the temperature and pressure parameters of the machine comply with the standards req...

Page 18: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 18 of 26 ELECTRICAL SYSTEM...

Page 19: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 19 of 26 HYDRAULIC PLANT...

Page 20: ...mostato automatico 1 25 3968 Porta tray 1 9 3937 Resistenza 500W 1 24 1730 Manometro 40 1 8 3862 Fusibile 20x4 F1A 1 22 0339 Micro interruttore 1 7 4787 Termostato bimetallico 1 21 2853 Pressostato 2...

Page 21: ...by fax to the number 0039 522 875579 and please wait the authorization for the shipment that must be with your agent at your charge 2 Unauthorized or not agreed shipments won t be accepted 3 Empty bo...

Page 22: ...with all the conditions ensuring that the appliance is correctly installed in compliance with sect 5 1 of the instruction manual The installer also declares that he has duly trained the user to opera...

Page 23: ...possessing manufacturing defects with the exclusion of expenses for Transfers Call rights Mileage reimbursement Shipment Replaced parts shall be returned free port The ordering party shall be charged...

Page 24: ...ATURE AND STAMP Via Don Luigi Sturzo 6 42021 BARCO DI BIBBIANO Reggio Emilia ITALY THIS PART OF THE CERTIFICATE MUST BE SENT BACK BY REGISTERED LETTER WITHIN 8 DAYS FROM THE DATE OF PURCHASE WARRANTY...

Page 25: ...TO WHOM THE SERIAL NUMBERS OF THE CONSIGNED AUTOCLAVES ARE ASSOCIATED IT MUST ALSO BE POSSIBLE TO TRACE THE AUTOCLAVES THE SERIAL NUMBER OF THE MACHINE MUST CORRESPOND TO THE END USER CUSTOMER EVEN WH...

Page 26: ...MACHINE POUR STERILISER A VAPEUR STEAM STERILIZER MACHINE ESTERILIZADORAS DE VAPOR DAMPFSTERILISATOREN Mod M9010 56 Art STERILIX BABY La Reverberi s r l dichiara sotto la propria responsabilit che i...

Reviews: