Operation and Maintenance Manual – Steam sterilizer mod.  Sterilix Baby 

Rev. 00 del  03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI 
ISTRUZIONE\Baby GB- R00.doc 

 

Pag. 11 of 26 

 

 

 

6. GENERAL INSTRUCTIONS FOR USE 

 

 

6.1  Method of use 

 

a)

 

Open the door by means of the device (10). 

b)

 

Turn on the main switch (7). 

c)

 

Arrange the instruments on the supplied Trays without overlapping the packets.  The maximum 
load allowed is 1 kg. 

d)

 

Fill the chamber with distilled water using the supplied dispenser, which holds 250 cc when full. 

e)

 

Shut the door using the handle (10). 

f)

 

Select the required sterilizing time using the timer (5). This time can vary depending on the load 
of instruments and the type of material.  

g)

 

Activate the selected sterilizing cycle with keys (2) or (3). 

h)

 

If very delicate materials are sterilized, it is advisable to remove them as soon as the cycle ends 
or even during the dry phase (1), after making sure that the pressure gauge (6) setting is Ø. 

i)

 

Each  time  that  a  cycle  is  carried  out,  place  an  indicator  (the  TERMALOG  or  VAPOR  LINE 
EXTENDER) in the sterilizing chamber to confirm the sterilizing process. 

j)

 

When the display (4) reaches Ø, sterilizing will have terminated. The door can then be opened 
with caution, always after having made sure that the pressure gauge (6) is on the Ø setting. 

k)

 

The machine switches off automatically at the end of the cycle and issues an acoustic signal. 

l)

 

If  the  instruments  are  still  wet  after  sterilizing,  proceed  as  follows  to  complete  the  drying 
process. 

 -leave the door ajar. 
 -select the Dry cycle (1). 
 -set the required number of minutes for the drying phase with the (5). 
 -now repeat the drying cycle, which will stop automatically. 

m)

 

Use the supplied tool and wear protective gloves when removing the material from the chamber. 

n)

 

Keep  about  50  cm  away  from  the  appliance  when  the  door  is  opened  as  residue  steam  could 
escape from the upper side of the door and could scald the operator. 

o)

 

Open the door with care, using the device (10) and an upward movement, always after checking 
that  the  pressure  gauge  (6)  setting  is  Ø.  If  no  resistance  is  opposed,  the  door  will  open  easily 
when the device (10) is moved from the right towards the left. 

p)

 

If the door is not opened immediately after the sterilizing cycle ends, condensed watery vapour 
could form on its surface and drip off on opening. 

 
 

6.2   End of work

 

 

After the operations have terminated, set the green main switch (10) to position 0, open the door and 
leave it ajar with handle (10) in the vertical position.. 

 

 

 
Pay particular attention to the temperature of the various components 
when you switch off the machine. 
 
Wait a few minutes to allow the components to cool.

 

 
 

 

Summary of Contents for M 9010/56

Page 1: ...Cod 4073 Operation and maintenance manual STEAM AUTOCLAVE MODEL STERILX BABY GB...

Page 2: ...d out by the technician when he installs the machine This form will be considered the reference certificate that validates the warranty It must be sent back to us along with the printer coupon and war...

Page 3: ...uge warning lights and by the position of the controls SYMBOLS AFFIXED TO THE MACHINE WARNING IDENTIFIES A HEAT SOURCE A PART COULD BE DANGEROUSLY HOT WARNING IDENTIFIES A GENERIC DANGER THERE COULD B...

Page 4: ...Five minute final drying phase Switch to select the 121 or 134 cycle Fully visible water filling from the top Built in drying cycle with evaporation system Continuous venting of the air in the chambe...

Page 5: ...ged by a skilled electrician The machines have been designed to operate up to an altitude of 1200 mm above sea level Contact the manufacturer if the machine must operate at a higher altitude as specia...

Page 6: ...ne to prevent accidental contact with hot parts Never ever operate the appliance unless the power cable is in a perfect condition and the structure undamaged It is advisable to keep the door half open...

Page 7: ...rticular preparations are required to operate the machines Just fit the plug into a 230 V power socket The power draw of the machines is 1500 Watt The control panel is elegantly designed easy to read...

Page 8: ...MANUAL 1 10 LITER CAN 1 250 CC DISPENSER DESCRIPTION OF THE REAR PART 12 Power cable with Shuko France plug 13 Fuse T10A 5x20 14 Regolation thermostat 15 Safety valve discharge 16 Used water drain tu...

Page 9: ...iately notify our dealer if discrepancies are discovered Keep the packing materials for as long as the machine remains in use not just for the warranty period This is because the autoclave may need to...

Page 10: ...ll the autoclave in a well ventilated place well away from sinks heat sources grinders and from all other machines that create vibrations and or dust Place the autoclave delicately on to a perfectly f...

Page 11: ...er having made sure that the pressure gauge 6 is on the setting k The machine switches off automatically at the end of the cycle and issues an acoustic signal l If the instruments are still wet after...

Page 12: ...ruments must be sterilized it is advisable to cover the trays with a paper or fabric napkin to prevent direct contact with the tray surface do not place the instruments straight on to the trays 8 Scis...

Page 13: ...1 2 2 1 2 2 2 1 2 2 1 1 1 2 Drying minutes 5 5 5 5 5 Sterilizing minutes with machine hot 23 33 18 23 23 Sterilizing minutes with machine cold 25 35 20 25 25 Max load Kg 1 1 1 1 0 5 Materials metallic...

Page 14: ...mply with the sterilizing parameters indicated by the current standards in force If the autoclave is used to sterilize instruments contaminated by CREUTZFELDT JAKOB s disease use the 134 cycle and ste...

Page 15: ...has tripped Press the reset alongside the control panel Switch 7 faulty Replace the switch Insufficient pressure for the selected cycle with indicator lights 2 and 3 permanently on The bulb thermostat...

Page 16: ...leave the door ajar and store it at a room temperature of 8 C Use a damp cloth to clean the casing of the machine Never ever use inflammable liquids for any reason whatsoever 10 2 Daily routine maint...

Page 17: ...autoclave to operate in a regular way the manufacturer advises users to carry out a functional test to ensure that the temperature and pressure parameters of the machine comply with the standards req...

Page 18: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 18 of 26 ELECTRICAL SYSTEM...

Page 19: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 19 of 26 HYDRAULIC PLANT...

Page 20: ...mostato automatico 1 25 3968 Porta tray 1 9 3937 Resistenza 500W 1 24 1730 Manometro 40 1 8 3862 Fusibile 20x4 F1A 1 22 0339 Micro interruttore 1 7 4787 Termostato bimetallico 1 21 2853 Pressostato 2...

Page 21: ...by fax to the number 0039 522 875579 and please wait the authorization for the shipment that must be with your agent at your charge 2 Unauthorized or not agreed shipments won t be accepted 3 Empty bo...

Page 22: ...with all the conditions ensuring that the appliance is correctly installed in compliance with sect 5 1 of the instruction manual The installer also declares that he has duly trained the user to opera...

Page 23: ...possessing manufacturing defects with the exclusion of expenses for Transfers Call rights Mileage reimbursement Shipment Replaced parts shall be returned free port The ordering party shall be charged...

Page 24: ...ATURE AND STAMP Via Don Luigi Sturzo 6 42021 BARCO DI BIBBIANO Reggio Emilia ITALY THIS PART OF THE CERTIFICATE MUST BE SENT BACK BY REGISTERED LETTER WITHIN 8 DAYS FROM THE DATE OF PURCHASE WARRANTY...

Page 25: ...TO WHOM THE SERIAL NUMBERS OF THE CONSIGNED AUTOCLAVES ARE ASSOCIATED IT MUST ALSO BE POSSIBLE TO TRACE THE AUTOCLAVES THE SERIAL NUMBER OF THE MACHINE MUST CORRESPOND TO THE END USER CUSTOMER EVEN WH...

Page 26: ...MACHINE POUR STERILISER A VAPEUR STEAM STERILIZER MACHINE ESTERILIZADORAS DE VAPOR DAMPFSTERILISATOREN Mod M9010 56 Art STERILIX BABY La Reverberi s r l dichiara sotto la propria responsabilit che i...

Reviews: