Operation and Maintenance Manual – Steam sterilizer mod.  Sterilix Baby 

Rev. 00 del  03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI 
ISTRUZIONE\Baby GB- R00.doc 

 

Pag. 12 of 26 

 

7.HOW TO ARRANGE THE INSTRUMENTS FOR THE 
STERILIZING PROCESS 

 

 

7.1   Standard sterilizing

 

Before  being  sterilized,  the  instruments  must  be  cleaned  to  remove  blood  residues,  incrustations, 
foreign bodies, etc. These impurities can damage the items being sterilized, the autoclave components 
and prevent the sterilizing process from being correctly carried out. 
These  instruments  must  be  made  of  NON-ferrous  material  since  they  could  damage  the  other 
instruments or the autoclave itself. 
To achieve perfect sterilizing, it is therefore important to proceed in the following way: 
1.  Rinse and brush the instruments under running water immediately after use to remove all deposits. 
2.  Place the instruments in the ultrasound machine with distilled water and detergent solution if 

possible. 

3.  Thoroughly rinse the instruments in demineralized water to prevent scaling stains. 
4.  Dry the instruments to eliminate all water residues which could lead to oxidation. 
5.  Wash, rinse and dry the supplied trays and make sure that they are perfectly clean. 
6.  To ensure that packaged instruments are perfectly dried, we recommend use of support art. M9050 

supplied as an optional. 

7.  If loose instruments must be sterilized, it is advisable to cover the trays with a paper or fabric 

napkin to prevent direct contact with the tray surface: do not place the instruments straight on to the 
trays. 

8.  Scissors and forceps should be slightly opened. Mirrors and containers should be placed upside 

down for the sterilizing process. 

9.  It is advisable to keep the instruments separated from each other to achieve a better sterilizing 

action. Overloading will impair the sterilizing process. 

10. If packaged instruments must be sterilized on the trays, place the packets with their transparent side 

downwards and do not overlap. The packets should not be too large. Drying will be difficult if their 
edges overlap. 

11. The machine can operate with a maximum 3 Kg load. 

 

7.2   Sterilzing handpieces

 

1.

 

Carefully read the handpiece operation manual with particular attention to the part about sterilizing. 

2.

 

Wrap the handpiece in the special sterilizing cloth. 

3.

 

Place the wrapped handpiece on a NON-perforated tray to prevent spilt lubricating oil from 
polluting the pipes of the autoclave. 

4.

 

Proceed with a short cycle 121°C sterilizing program. 

5.

 

Remember to immediately remove the handpiece from the autoclave at the end of the sterilizing 
process. 

6.

 

It is inadvisable to sterilize handpieces using the night cycle. 

It is essential to place a chemical indicator in the autoclave on each cycle, to ensure that the sterilizing 
process has taken place correctly. 

 

7.3 Messaggi visualizzati dal display 

 

P1 

Cycle 121° 

P2 

Cycle 134° 

dr 

Dry cycle 

IIe 

Heating 

St 

Steril 

 
 

Summary of Contents for M 9010/56

Page 1: ...Cod 4073 Operation and maintenance manual STEAM AUTOCLAVE MODEL STERILX BABY GB...

Page 2: ...d out by the technician when he installs the machine This form will be considered the reference certificate that validates the warranty It must be sent back to us along with the printer coupon and war...

Page 3: ...uge warning lights and by the position of the controls SYMBOLS AFFIXED TO THE MACHINE WARNING IDENTIFIES A HEAT SOURCE A PART COULD BE DANGEROUSLY HOT WARNING IDENTIFIES A GENERIC DANGER THERE COULD B...

Page 4: ...Five minute final drying phase Switch to select the 121 or 134 cycle Fully visible water filling from the top Built in drying cycle with evaporation system Continuous venting of the air in the chambe...

Page 5: ...ged by a skilled electrician The machines have been designed to operate up to an altitude of 1200 mm above sea level Contact the manufacturer if the machine must operate at a higher altitude as specia...

Page 6: ...ne to prevent accidental contact with hot parts Never ever operate the appliance unless the power cable is in a perfect condition and the structure undamaged It is advisable to keep the door half open...

Page 7: ...rticular preparations are required to operate the machines Just fit the plug into a 230 V power socket The power draw of the machines is 1500 Watt The control panel is elegantly designed easy to read...

Page 8: ...MANUAL 1 10 LITER CAN 1 250 CC DISPENSER DESCRIPTION OF THE REAR PART 12 Power cable with Shuko France plug 13 Fuse T10A 5x20 14 Regolation thermostat 15 Safety valve discharge 16 Used water drain tu...

Page 9: ...iately notify our dealer if discrepancies are discovered Keep the packing materials for as long as the machine remains in use not just for the warranty period This is because the autoclave may need to...

Page 10: ...ll the autoclave in a well ventilated place well away from sinks heat sources grinders and from all other machines that create vibrations and or dust Place the autoclave delicately on to a perfectly f...

Page 11: ...er having made sure that the pressure gauge 6 is on the setting k The machine switches off automatically at the end of the cycle and issues an acoustic signal l If the instruments are still wet after...

Page 12: ...ruments must be sterilized it is advisable to cover the trays with a paper or fabric napkin to prevent direct contact with the tray surface do not place the instruments straight on to the trays 8 Scis...

Page 13: ...1 2 2 1 2 2 2 1 2 2 1 1 1 2 Drying minutes 5 5 5 5 5 Sterilizing minutes with machine hot 23 33 18 23 23 Sterilizing minutes with machine cold 25 35 20 25 25 Max load Kg 1 1 1 1 0 5 Materials metallic...

Page 14: ...mply with the sterilizing parameters indicated by the current standards in force If the autoclave is used to sterilize instruments contaminated by CREUTZFELDT JAKOB s disease use the 134 cycle and ste...

Page 15: ...has tripped Press the reset alongside the control panel Switch 7 faulty Replace the switch Insufficient pressure for the selected cycle with indicator lights 2 and 3 permanently on The bulb thermostat...

Page 16: ...leave the door ajar and store it at a room temperature of 8 C Use a damp cloth to clean the casing of the machine Never ever use inflammable liquids for any reason whatsoever 10 2 Daily routine maint...

Page 17: ...autoclave to operate in a regular way the manufacturer advises users to carry out a functional test to ensure that the temperature and pressure parameters of the machine comply with the standards req...

Page 18: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 18 of 26 ELECTRICAL SYSTEM...

Page 19: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 19 of 26 HYDRAULIC PLANT...

Page 20: ...mostato automatico 1 25 3968 Porta tray 1 9 3937 Resistenza 500W 1 24 1730 Manometro 40 1 8 3862 Fusibile 20x4 F1A 1 22 0339 Micro interruttore 1 7 4787 Termostato bimetallico 1 21 2853 Pressostato 2...

Page 21: ...by fax to the number 0039 522 875579 and please wait the authorization for the shipment that must be with your agent at your charge 2 Unauthorized or not agreed shipments won t be accepted 3 Empty bo...

Page 22: ...with all the conditions ensuring that the appliance is correctly installed in compliance with sect 5 1 of the instruction manual The installer also declares that he has duly trained the user to opera...

Page 23: ...possessing manufacturing defects with the exclusion of expenses for Transfers Call rights Mileage reimbursement Shipment Replaced parts shall be returned free port The ordering party shall be charged...

Page 24: ...ATURE AND STAMP Via Don Luigi Sturzo 6 42021 BARCO DI BIBBIANO Reggio Emilia ITALY THIS PART OF THE CERTIFICATE MUST BE SENT BACK BY REGISTERED LETTER WITHIN 8 DAYS FROM THE DATE OF PURCHASE WARRANTY...

Page 25: ...TO WHOM THE SERIAL NUMBERS OF THE CONSIGNED AUTOCLAVES ARE ASSOCIATED IT MUST ALSO BE POSSIBLE TO TRACE THE AUTOCLAVES THE SERIAL NUMBER OF THE MACHINE MUST CORRESPOND TO THE END USER CUSTOMER EVEN WH...

Page 26: ...MACHINE POUR STERILISER A VAPEUR STEAM STERILIZER MACHINE ESTERILIZADORAS DE VAPOR DAMPFSTERILISATOREN Mod M9010 56 Art STERILIX BABY La Reverberi s r l dichiara sotto la propria responsabilit che i...

Reviews:

Related manuals for M 9010/56