Operation and Maintenance Manual – Steam sterilizer mod. Sterilix Baby
Rev. 00 del 03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI
ISTRUZIONE\Baby GB- R00.doc
Pag. 14 of 26
* Until he has acquired enough experience in using the autoclave, the operator must check to make sure
that the operating pressure does not last less than the minimum effective sterilization minutes indicated
in the table.
Increase the number of minutes that sterilizing must last if these parameters are not complied with.
The sterilizing table contains indicative data. It is essential to consult the values established by the
manufacturer of the materials to be sterilized and to strictly comply with the sterilizing parameters
indicated by the current standards in force.
If the autoclave is used to sterilize instruments contaminated by CREUTZFELDT-JAKOB's disease,
use the 134° cycle and sterilize for 20 minutes with a cold start or 20 minutes with a hot start.
It is advisable to use non-packeted sterilized materials immediately. Packeted sterilized instruments can
be placed in a UV glass case to ensure that they remain sterilized (for short periods).
8.2 Notes and warnings
1- Strictly comply with the manufacturer's instructions when sterilizing materials.
2- To preserve packaged and loose sterilized materials, the trays must be stored in special cabinets
fitted with ultraviolet lamps and designed for this specific use.
3- Whenever a sterilizing process is carried out, it is advisable to add a chemical indicator that we can
supply on request, code 1468.