Operation and Maintenance Manual – Steam sterilizer mod.  Sterilix Baby 

Rev. 00 del  03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI 
ISTRUZIONE\Baby GB- R00.doc 

 

Pag. 16 of 26 

 

 

 

 10. MAINTENANCE 

 

 

 

 

 

Carry out the maintenance operations with the machine at a standstill, after having 
removed the plug from the power socket and after having allowed the components to 
cool down. 
The  declaration  of  conformity  to  the  standards  in  force  is  not  valid  unless  the 
maintenance operations described below are carried out.

 

 

10.1 General maintenance 

 
The machine needs to be serviced in a particular way. Users are therefore advised to comply with the 
below listed recommendations: 
 



Periodically check the condition of the electrical system with particular attention to the connection 
cable. 

Contact  our  Technical  Assistance  Service  or  specialized  personnel  if  faults  are 

discovered.

 



Periodically clean the door, the seal and take particular care when cleaning the 

internal part of the 

boiler

, removing any scaling with the damp sponge (supplied with the machine). Correct cleaning 

of these parts of the machine will ensure that the sterilizing cycle takes place in the right way. 



If the machine is not to be used for a long period of time, leave the door ajar and store it at a room 
temperature of >8°C. 



Use a damp cloth to clean the casing of the machine. Never ever use inflammable liquids for any 
reason whatsoever. 

 
 

10.2 Daily routine maintenance 

 
Clear the door seal, door window and generally clean the outer and inner surfaces of the machine. 
Clean the black silicone door seal, the outer and inner edges and the part where the door rests on the 
sterilizing chamber and on which the seal must tightly fit. Use the supplied sponge to do this. The soft 
part of the sponge can be used to clean the seal while the rough part can be used to clean the edge of the 
chamber. 
This cleaning operation must be carried out regularly to remove all impurities that could lead to loss of 
pressure inside the sterilizing chamber. 
Check the used water level and drain off the actual water into the sewer if necessary. Then wash out the 
can under running water. 
 

 

10.3 Weekly routine maintenance 

 
Cleaning the sterilizing chamber / cleaning the trays and tray holders. 
As specified in section 10.1 “ General maintenance “, all traces of deposits must be removed from the 
bottom  of  the  chamber.  Use  the  rough  side  of  the  supplied  sponge  for  this  operation,  attempting  to 
eliminate any scaling that may have formed. 
Rinse everything with the same water used for the sterilizing process. The trays and tray-holders must 
also be cleaned in this way. 
Drain the water and clean the tray. 
 

Summary of Contents for M 9010/56

Page 1: ...Cod 4073 Operation and maintenance manual STEAM AUTOCLAVE MODEL STERILX BABY GB...

Page 2: ...d out by the technician when he installs the machine This form will be considered the reference certificate that validates the warranty It must be sent back to us along with the printer coupon and war...

Page 3: ...uge warning lights and by the position of the controls SYMBOLS AFFIXED TO THE MACHINE WARNING IDENTIFIES A HEAT SOURCE A PART COULD BE DANGEROUSLY HOT WARNING IDENTIFIES A GENERIC DANGER THERE COULD B...

Page 4: ...Five minute final drying phase Switch to select the 121 or 134 cycle Fully visible water filling from the top Built in drying cycle with evaporation system Continuous venting of the air in the chambe...

Page 5: ...ged by a skilled electrician The machines have been designed to operate up to an altitude of 1200 mm above sea level Contact the manufacturer if the machine must operate at a higher altitude as specia...

Page 6: ...ne to prevent accidental contact with hot parts Never ever operate the appliance unless the power cable is in a perfect condition and the structure undamaged It is advisable to keep the door half open...

Page 7: ...rticular preparations are required to operate the machines Just fit the plug into a 230 V power socket The power draw of the machines is 1500 Watt The control panel is elegantly designed easy to read...

Page 8: ...MANUAL 1 10 LITER CAN 1 250 CC DISPENSER DESCRIPTION OF THE REAR PART 12 Power cable with Shuko France plug 13 Fuse T10A 5x20 14 Regolation thermostat 15 Safety valve discharge 16 Used water drain tu...

Page 9: ...iately notify our dealer if discrepancies are discovered Keep the packing materials for as long as the machine remains in use not just for the warranty period This is because the autoclave may need to...

Page 10: ...ll the autoclave in a well ventilated place well away from sinks heat sources grinders and from all other machines that create vibrations and or dust Place the autoclave delicately on to a perfectly f...

Page 11: ...er having made sure that the pressure gauge 6 is on the setting k The machine switches off automatically at the end of the cycle and issues an acoustic signal l If the instruments are still wet after...

Page 12: ...ruments must be sterilized it is advisable to cover the trays with a paper or fabric napkin to prevent direct contact with the tray surface do not place the instruments straight on to the trays 8 Scis...

Page 13: ...1 2 2 1 2 2 2 1 2 2 1 1 1 2 Drying minutes 5 5 5 5 5 Sterilizing minutes with machine hot 23 33 18 23 23 Sterilizing minutes with machine cold 25 35 20 25 25 Max load Kg 1 1 1 1 0 5 Materials metallic...

Page 14: ...mply with the sterilizing parameters indicated by the current standards in force If the autoclave is used to sterilize instruments contaminated by CREUTZFELDT JAKOB s disease use the 134 cycle and ste...

Page 15: ...has tripped Press the reset alongside the control panel Switch 7 faulty Replace the switch Insufficient pressure for the selected cycle with indicator lights 2 and 3 permanently on The bulb thermostat...

Page 16: ...leave the door ajar and store it at a room temperature of 8 C Use a damp cloth to clean the casing of the machine Never ever use inflammable liquids for any reason whatsoever 10 2 Daily routine maint...

Page 17: ...autoclave to operate in a regular way the manufacturer advises users to carry out a functional test to ensure that the temperature and pressure parameters of the machine comply with the standards req...

Page 18: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 18 of 26 ELECTRICAL SYSTEM...

Page 19: ...Operation and Maintenance Manual Steam sterilizer mod Sterilix Baby Rev 00 del 03 01 2007 SERVER2012 oo SETTORE DENTALE BABY MANUALI ISTRUZIONE Baby GB R00 doc Pag 19 of 26 HYDRAULIC PLANT...

Page 20: ...mostato automatico 1 25 3968 Porta tray 1 9 3937 Resistenza 500W 1 24 1730 Manometro 40 1 8 3862 Fusibile 20x4 F1A 1 22 0339 Micro interruttore 1 7 4787 Termostato bimetallico 1 21 2853 Pressostato 2...

Page 21: ...by fax to the number 0039 522 875579 and please wait the authorization for the shipment that must be with your agent at your charge 2 Unauthorized or not agreed shipments won t be accepted 3 Empty bo...

Page 22: ...with all the conditions ensuring that the appliance is correctly installed in compliance with sect 5 1 of the instruction manual The installer also declares that he has duly trained the user to opera...

Page 23: ...possessing manufacturing defects with the exclusion of expenses for Transfers Call rights Mileage reimbursement Shipment Replaced parts shall be returned free port The ordering party shall be charged...

Page 24: ...ATURE AND STAMP Via Don Luigi Sturzo 6 42021 BARCO DI BIBBIANO Reggio Emilia ITALY THIS PART OF THE CERTIFICATE MUST BE SENT BACK BY REGISTERED LETTER WITHIN 8 DAYS FROM THE DATE OF PURCHASE WARRANTY...

Page 25: ...TO WHOM THE SERIAL NUMBERS OF THE CONSIGNED AUTOCLAVES ARE ASSOCIATED IT MUST ALSO BE POSSIBLE TO TRACE THE AUTOCLAVES THE SERIAL NUMBER OF THE MACHINE MUST CORRESPOND TO THE END USER CUSTOMER EVEN WH...

Page 26: ...MACHINE POUR STERILISER A VAPEUR STEAM STERILIZER MACHINE ESTERILIZADORAS DE VAPOR DAMPFSTERILISATOREN Mod M9010 56 Art STERILIX BABY La Reverberi s r l dichiara sotto la propria responsabilit che i...

Reviews:

Related manuals for M 9010/56