Operation and Maintenance Manual – Steam sterilizer mod. Sterilix Baby
Rev. 00 del 03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI
ISTRUZIONE\Baby GB- R00.doc
Pag. 4 of 26
2. TECHNICAL SPECIFICATIONS
This autoclave
as been designed and built in compliance with the operators' need for functionality,
practical use and safety.
This autoclave
has functional characteristics that make it unique, i.e.:
Main switch to power the machine.
60 minute cycle program.
Five minute final drying phase.
Switch to select the 121° or 134° cycle.
Fully visible water filling from the top.
Built in drying cycle with evaporation system.
Continuous venting of the air in the chamber to encourage the maximum steam saturation.
Approved safety valves to prevent over-pressures.
Safety valve to protect the heating element.
Safety thermostat to monitor the drying operation.
121° - 134° operating pressure switches.
Chamber in seamless press-moulded 18/10 STAINLESS STEEL.
7 liter chamber capacity able to hold up to 3 trays.
Condensation radiator to drain off the water at the end of the cycle.
Used water can
Filling with the 250 cc dispenser.