Operation and Maintenance Manual – Steam sterilizer mod. Sterilix Baby
Rev. 00 del 03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI
ISTRUZIONE\Baby GB- R00.doc
Pag. 5 of 26
Description
Unit of
measurement
Values
Power source voltage rating
V
Alternata 230 ~
Power source frequency
Hz
50/60
Machine power rating
W
1000
Boiler volume
mm
Ø 170 x 270
Operating pressure
Bar
2.1 - 1.1
Machine dimensions X, Y, Z
mm
390 x 370 x 300
Gross weight
Kg
20
Pack dimensions
mm
470 x 450 x 380
Operating temperatures
°C
+ 3°C
……
+ 35°C
Distance from walls
mm
20
X = depth
Y = width
Z = height
The electrical system is protected against the risk of over-currents by two 10 A
6,3x32 delayed action fuses ( T 10 A ) installed inside the machine.
Only have these fuses changed by a skilled electrician.
The machines have been designed to operate up to an altitude of 1200 mm above sea
level. Contact the manufacturer if the machine must operate at a higher altitude as
special models and/or devices may be required.
The noise level of the machine has been tested. The continuous equivalent
weighted acoustic pressure level A is less than 70 dB (A) and the continuous
equivalent weighted acoustic power level A does not exceed 85 dB (A). The
measured noise level measured is therefore very low. For this reason, the operator
need not take any particular measures (also in view of the fact that the machine
operates automatically).
DATA AND SPECIFICATIONS OF THE PRESSURIZED VESSEL:
Serial N° of vessel, Serial N° of door
Month N°, Year N° (see internal label)
Max. operating pressure:
2,2 Bar.
Max. operating temperature:
160°C
Min. operating pressure:
-0,9 Bar.
Test pressure:
8 Bar.
Reference standards:
TRD 421
Volume of vessel:
7 lt.