Operation and Maintenance Manual – Steam sterilizer mod. Sterilix Baby
Rev. 00 del 03-01-2007 - \\SERVER2012\oo\SETTORE DENTALE\BABY\MANUALI
ISTRUZIONE\Baby GB- R00.doc
Pag. 9 of 26
5. STARTING AND SETTING AT WORK
5.1 Removing the machine from the packing
The machine is consigned packed on a wooden pallet out a cardboard box: make sure that there are no
dents and that the wrapping is in optimum conditions.
The machine must be stored in a dry place, sheltered from adverse weather conditions, at a temperature
b3°C and +35°C.
The machine must be handled without jolting and using appropriate means.
Open the pack without overturning it.
Remove the polystyrene from the top and sides (the weight of the model is given in chap. 2, technical
specifications).
Remove the pluriball wrapping from around the machine and make sure that all parts of this latter are
undamaged and complete with all the accessories indicated on the consignment note. Immediately notify
our dealer if discrepancies are discovered.
Keep the packing materials for as long as the machine remains in use, not just for the warranty period.
This is because the autoclave may need to be calibrated or regulated by the manufacturer's staff of by an
after-sales center. Non-original packing is unable to ensure that the machine is correctly protected during
transport.
The original packing comprises the following items:
- BIWALL BOX IN CARDBOARD WITH 3 CORRUGATIONS
- PLURIBALL POLYETHYLENE BAG
- SHAPED POLYSTYRENE FOAM
- PALLET OUTSIDE THE BOX
CEN STANDARD DIN EN 285
Maximum value
Evaporation residue
10 mg/l
Silicon dioxide (SiO
2
)
1 mg/l
Iron
0.2 mg/l
Cadmium
0.005 mg/l
Lead
0.05 mg/l
Heavy metal residues (barring iron, cadmium, lead)
0.1 mg/l
Chloride
2 mg/l
Phosphate
0.5 mg/l
Conductivity (at 20°C)
15 ųs/cm
pH value
From 5 to 7
Appearance
Colourless, clean and sediment free
Hardness
0.02 mmol/l
NOTE: The use of water containing higher concentrations than those indicated in the table above could
considerably reduce the life of the equipment, causing serious damages to its components and
consequently annulling the warranty.