SERENA Tecnico GB R-06 del 10.05.2019
00/ Per Bertani/ Manuale Tecnico Serena/ Serena
5
1.3 CALIBRATION
Warning! Before proceeding to calibration, the machine must be
completely cold and preheating must be switched off!
LOGO
SERIAL NUMBER
SW VERSION
CALIBRATION
INCREASE BAR
FILLING
The following screen will appear
CHAMBER PR.
T1 CHAMBER
T2 CHAMBER
T3 VAPORIZER
T4 BODY
T5 RESISTANCE
By using a certified instrument, adjust the calibration so that the data on the display corresponds to the test
instrument.
1.3.1 CHAMBER PRESSURE
CHAMBER PR.
+000 mbar
1000 mbar