Operation and Maintenance Manual – Steam sterilizers mod. PRATIKA
Rev. 01 del 03-01-2007 - C:\Documenti\Manuali di istruzione\PRATIKA GB-R01.doc
Pag. 15 di 27
Sterilizing handpieces:
Carefully read the handpiece operation manual with particular attention to the part about sterilizing
(121°).
Wrap the handpiece in the special sterilizing cloth.
Place the wrapped handpiece on a NON-perforated tray to prevent spilt lubricating oil from
polluting the pipes of the autoclave.
Proceed with a short cycle 121°C sterilizing program.
Remember to immediately remove the handpiece from the autoclave at the end of the sterilizing
process.
For a valid sterilizing process, it is obligatory to add a chemical indicator for each cycle.
7.1 Sterilizing table
CYCLES FREE
STERILIZING TIME (Min.)
DRYING TIME (Min.)
Nr.
PICCHI
1 *
*
2 *
*
1-3
CYCLES AT 121°
STERILIZING TIME (Min.)
DRYING TIME (Min.)
Nr.
PICCHI
3 UNWRAPPED
15
5
4 WRAPPED
20
15
1
CYCLES AT 134°
STERILIZING TIME (Min.)
DRYING TIME (Min.)
Nr.
PICCHI
5 UNWRAPPED
4
5
6 WRAPPED
10
15
7 PRIONE
18
20
1
TEST
STERILIZING TIME (Min.)
DRYING TIME (Min.)
Nr.
PICCHI
BOWIE & DICK*
3,5
2
Vacuum**
15 (Total length of test)
1
CYCLE PHIAL
STERILIZING TIME (Min.)
DRYING TIME (Min.)
54°
48h (Total length of test)
* Preset value. ** To execute with cold machine.
7.2 Notes and warnings
1- Strictly comply with the manufacturer's instructions when sterilizing materials.
2- To preserve packaged and loose sterilized materials, the trays must be stored in special cabinets
fitted with ultraviolet lamps and designed for this specific use.
3- If the articles are to be stored for a longer period, it is advisable to use our covered trays code 1474 +
2682 and tray holders code 2631 supplied on request.
4- Whenever a sterilizing process is carried out, it is advisable to add a chemical indicator that we can
supply on request, code 1468.