SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
11/99
warranty remains unchanged: it always starts from the original date of purchase of the unit.
As nothere reported The Terms and Conditions listed on the commercial documentation / sales catalogue in force
at the time of purchase remain valid.
1.8 ENVIRONMENTAL CONDITIONS ALLOWED
TEMPERATURE, HUMIDITY AND ALTITUDE
See Paragraph in particular
ELECTROMAGNETIC COMPATIBILITY
Install the unit where electromagnetic interference comply with the limits permitted by the rules and with the presence
of electric and electronic equipment.
RADIATION EXPOSURE
The appliance is not intended to be used to direct exposure to:
- Non-ionizing radiation of the sun, being the same to be placed indoor and / or repaired
- X-
ray radiation, microwave radiation, laser radiation, etc…
In the event that, in the environment where the unit is placed, there is a risk of radiation directly on it, specific measures
to protect exposed areas must be taken by the purchaser.
USE IN EXPLOSIVE ENVIRONMENT
The device cannot be used in any way in the environment where there is a risk of explosion or in a potentially explosive
environment.
IT IS FORBIDDEN TO USE THE EQUIPMENT IN EXPLOSIVE
ENVIRONMENTS SUCH OR CONSIDERED.
1.9 ILLUMINATION IN THE WORKPLACE
The environment in which the sterilizer is installed must meet the requirements listed on the Directive 89/654/EEC
concerning the workplace and the requirements of the RULES FORCE THE INSTALLATION SITE.
Proper lighting can, in some cases, avoid or limit an injury to the operator.
REQUIREMENTS OF THE LOCAL WHERE THE EQUIPMENT IS INSTALLED. THE LOCAL SHOULD:
• be free of debris and / or residues of various kinds of flammable or potentially dangerous operation;
• provide a specific room for sterilization, away from the patient.
• have doors (accesses) appropriate to the size of the machine, ensure a suitable floor for the mass flow of the
machine, ensure a perfectly level floor, be free, at the area of installation, strong drafts or substantial temperature
changes;
• ensure good ventilation and air exchange with the provision of specific natural ventilation openings.