4.2
CONNECTION TO THE LABELLER M9006 (ON DEMAND)
Use only our labeller UNION PRINT!
The use of other labellers relieves the manufacturer of the autoclave from any liability
under warranty or any other claim or damage caused same label.
-
Use only thermal labels supplied by us - Place the roll as shown in the drawing visible when you open the door - Close the door until it
clicks leaving a label outside - Press the button to place the label.
-
Flashing red light -> NOT operational
-
Flashing green light -> WAITING positioning
-
Solid green light -> READY
To do a test print (p. 44 Instruction Manual): OPTIONS -> LABEL PRINTING LAST CYCLE
In the case of incorrect printing, too high or too low, the light is solid green, press and hold the button and wait until the light flashes once and
then twice more, then release the button
.
Use only original rolls and place them in the right way!
4.3
CONNECTIONS TO DEMINERALIZERS KIARA-TEC (ON DEMAND)
The manufacturer offers the possibility of automatic feeding of SERENA autoclaves with demineralized water. Once
installed the system, the tank of clean water is automatically kept full without any operator intervention.
4.4
CONTINUOUS DRAIN CONNECTION (ON DEMAND)
The autoclave can be connected to an external drain to allow the automatic discharge of water used.
The length the drain hose must not exceed 5 meters!
The discharge point must be at least 20-30 cm below the plane of the autoclave!