LOADING PREPARATION
5.0 CLEANING OF TOOLS
The objects to be sterilized must be perfectly clean, carefully dried and without any residue of blood, soap, plaster.
These substances can cause irremediable damage to both the objects to be sterilized and either to the components of
the autoclave as well as compromising the autoclave sterilization. In cleaning, always use protective gloves.
1. Immediately after the use, immerse the instruments in a bowl with disinfectant solution. Strictly follow what is
indicated by the manufacturer of the disinfectantfor the times and the percentages of solution. After this
preliminary decontamination proceed with a thorough rinsing.
2. Continue with cleansing by using ultrasound machines and washer machine. If you use ultrasound machines, it
is essential a thorough rinsing with demineralised water if possible. Follow the manufacturer's instructions. We
recommend using a WASHER-DISINFECTOR and disinfection that in addition to cleaning, rinse and dry
automatically.
3. Before the sterilization, make sure that the instruments are perfectly dry (use a paper towel, TNT or cloth of low
release) and without any residual disinfectant, since these residues may cause corrosion during the sterilization.
For the drying of hollow instruments use a compressed air gun.
4. Make a final check of the tools and eliminate those with cracks, stains, rust and friction between the parties (it
may be useful a magnifying glass). Then proceed with the enveloping.
5. For dynamic tools, lubricate according to the instructions provided by the manufacturer of the instruments.
6. Snippy or cutting materials should be protected with rubber tubes resistant to high temperatures.
7. For the cleaning of dental manipulates, proceed as follows:
•
Always use protective gloves
• Follow the instructions provided by the manufacturer of the tools
• Do not use abrasive cleaners nor chlorine-based disinfectants neither similar
•Do not immerse instruments in the disinfectant
• Clean, rinse and dry with extreme care, taking care not to scratch the fibre optic
• Lubricate the instrument before the sterilization according to the manufacturer's instructions
• Wrap the instrument in
Do not use sharp objects for cleaning or Disinfectants
Carefully clean the bottom of the chamber, taking care not to damage the sensors
Use non-abrasive cloth and light detergent.
Rinse thoroughly
WARNING! DO NOT introduce sensitive materials in the autoclave !