SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
46/99
7.2 EXTRA DRYING
With this menu the standard drying sterilization cycle is increased from 1 to 99 minutes.
Pressing the "START" you will have access to change information. This drying is indicated in the presence of multiple
packaging (double envelope). Continuing with the arrows you can access the other sub-menu OPTIONS
EXTRA DRYING
01 min
Press ESC to exit and return to the main menu
OPTIONS
LAST CYCLES
DRYING ONLY
EXTRA DRYING
LABELS EXPIRY
DATE
7.3 LABELS EXPIRY DATE
Pressing the "START" you will have access to change information. And 'the expiry date printed on the labels has to be
attached to the envelopes containing the sterilized instruments. After this date the instruments is no longer considered
sterilized.
LABELS EXPIRY
DATE
60
Press “ESC” to exit
OPTIONS
EXTRA DRYING
LABELS EXPIRY
DATE
START POSTPONED
CYCLE
7.4 START POSTPONED CYCLE
Pressing the "down arrow-up arrow" you should program a delayed start of the sterilization cycle up to 99 hours.
START POSTPONED
CYCLE
99
T
he next screen will appear all the cycles. Select the cycle and close the door.