12 SOLUTION TO FAILURES
PROBLEM
CAUSES
ACTION
The autoclave does not turn on
The main switch is off
No power outlet
Faulty cable
Turn on the power switch
Check the installation
Replace the cable
Loss of water in front of the
machine during the cycle
Gasket worn
Replace the gasket
Presence of water within the
chamber at the end of the
sterilization cycle
Problem inside the
machine
Contact authorized service
centre
Sterilized instruments with spots
or rust
Tools are not perfectly dry
Chemical residues on the
tools
Tools are not perfectly
clean
Tools made of different
materials wrapped
together
Sterilization chamber not
clean
Load water of poor quality
- see paragraph "water
quality"
Cutters or similar tools not
properly wrapped
Dry thoroughly the tools
before the sterilization
Rinse carefully the tools
Clean tools carefully
Place on several trays the
tools of different material
Always keep the chamber
clean - see paragraph
"cleaning and
maintenance"
Use water with features
referred to paragraph
"water quality"
See paragraph "load
preparation"
Tools that change colour
Choice of the wrong cycle
Tools fallen in the bottom
of the chamber
Always follow the tools
manufacturer's instructions
Ensure the correct
conditions to restore load
water
The display does not light
Fuse no. 1 burned
Flat cable not connected
Replace the fuse (identify
the cause)
Connect the flat cable
The door will not open.
The device is delivered
with the locked door
Options
Opening door
If the door does not open, contact
the service center.
Error message on the display
Operating error
Press the up arrow and
down arrow keys
simultaneously for at least
20 seconds
To break the cycle
Various reasons
Press the start button until
it beeps