SERENA P GB-R04 del 03-07-17
00/Per Bertani/Manuali d’uso Serena/ Serena con stampante
75/99
INTERRUPTION OF THE STERILIZATION CYCLE
VOLUNTARY INTERRUPTION
during the depression phase without introduction of water in the chamber:
1. Acoustic signal with the writing on the display CYCLE INTERRUPTED
2. Opening of the valve of the bacterial filter (no.7). The sterilizer comes to ambient pressure.
3. Starting of the vacuum pump (No.3) and electrovalve drain ignition (no.5) for a few minutes.
4. Opening of the three-way valve (number 6) up to 600 mbar
5. The sterilizer stops and the bacterial filter is activated .At ambient pressure, the door opens automatically and the
display will show the message CYCLE INTERRUPTED - LOAD NOT STERILE
6. Press START to remove the warning message.
LACK OF POWER TO THE STERILIZER:
1. Acoustic signal with a message on the display showing ALARM a0002 - LOAD NOT STERILE
2. Opening of the valve of the bacterial filter (no.7). The sterilizer reaches the ambient pressure.
3. Starting of the vacuum pump (No.3) and electrovalve drain ignition (no.5) for a few minutes.
4. Opening of the three-way valve (number 6) up to 600 mbar.
5. The sterilizer stops and the bacterial filter is activated. To open the door go in the MENU, OPTIONS and select OPEN
DOOR - START -START.
6. The display will show PLEASE WAIT. After a few seconds the door will open.
7. In the event that the sterilizer fails to fall below 600 mbar, e.g. for a pump problem, after 4 minutes the door will open
automatically, because the water inside the chamber will already have been eliminated.
VOLUNTARY INTERRUPTION during the SPECIALcycle1
1. Acoustic signal with a message on the display showing ALARM CYCLE INTERRUPTED - PLEASE WAIT
2. See points as described above
3. In the event the power failures to work while the sterilizer is being emptied (PLEASE WAIT) and returns the
current, the machine returns to the MENU '. It will open the door manually as in step 5.
VOLUNTARY INTERRUPTION during the SPECIALcycle1 duringthe sterilization phase
1. Acoustic signal with a message on the display showing ALARM INTERRUPTED CYCLE-PLEASE WAIT
2. Opening of the valve (no.5) and valve (No.4). the fan starts working. (no. 19).
3. Discharge water up to 1100 mbar
4. Starting of the vacuum pump (No.3) and opening of the valve (number 6) with the creation of vacuum inside the
chamber for a few minutes
5. On the display-CYCLE INTERRUPTED LOAD NOT STERILE
6. Door opens automatically when achieving 600mbar.
VOLUNTARY INTERRUPTION during the SPECIALcycle1 during the drying phase
1. Acoustic signal with a message showing on the display CYCLE INTERRUPTED
2. Opening the valve of the bacterial filter (no.7). The sterilizer comes to ambient pressure.
3. Starting of the vacuum pump (No.3) and ignition of the electrovalve drain (no.5) for a few minutes. The fan starts
working nr.19).
4. Opening the three-way valve (number 6) up to 60 mbar.
5. The sterilization stops and the bacteriological filter starts working. At ambient pressure, the door opens
automatically and the display will read-CYCLE INTERRUPTED LOAD NOT STERILE
6. Press START to remove the warning message.
NOTE: ALL TRANSACTIONS DESCRIBED ABOVE HAVE BEEN STUDIED TO PREVENT RELEASE OF WATER FROM THE
STERILIZATION CHAMBER.