SERENA P GB-R04 del 03-07-17
00/ Per Bertani/ Manuali d’uso Serena/ Serena con stampante
86/99
Form 93/A
FOR THE AUTHORIZED TECHNICIAN
SPARE PARTS KIT
Serena
SCHEDULED MAINTENANCE WITHIN TWO YEARS OR 2000 CYCLES
Coupon on Payment:
WITHI
N
20
00
CYCLES
Quantity
Code
Components description
REPLACED
REVISED
CONTROLLED
4 per 5104
6 per 5237
KNF118279
Gasket vacuum pump
1
5695
Vaporizer group
1
1
5104
5237
Vacuum pump 18 lt.
Vacuum pump 23 lt.
1
3028
Discharge electro valve
1
3246
3-ways electro valve
1
3028
Purifier electro valve
1
5158
Door micro switch
1
5040
Door pin
Registration and control door pressure
Replacement water filter
Cleaning tanks
Lubrication door plug
Control actuators
Control temperature parameters / pressure and test cycles
Control metallic filter inside the chamber
Electrical safety test
Replacement of
SAFETY THERMOSTATS
of the heaters
Check if the daily maintenance are performed, such as:
Check the water quality
Cleaning the door, the gasket and the sterilization chamber
Cleaning the exterior of the sterilizer
Send a copy of the revised delivery note to the manufacturer.