SERENA Tecnico GB R-04 del 04.04.17
00/ Per Bertani / Ma nuale Tecnico Serena / Serena
Pagina 5
1.3 CALIBRATION
Warning! Before proceeding to calibration, the machine must be completely
cold and preheating must be switched off!
LOGO
SERIAL NUMBER
SW VERSION
CALIBRATION
INCREASE BAR
FILLING
The following screen will appear
CHAMBER PR.
T1 CHAMBER
T2 CHAMBER
T3 VAPORIZER
T4 BODY
T5 RESISTANCE
By using a certified instrument, adjust the calibration so that the data on the display corresponds to the test
instrument.
1.3.1 CHAMBER PRESSURE
CHAMBER PR.
-004 mbar
1000 mbar
1.3.2 T1 CHAMBER
– TEMPERATURE PROBE
CHAMBER PR.
T1 CHAMBER
T2 CHAMBER
T3 VAPORIZER
T4 BODY
T5 RESISTANCE
T1 CHAMBER
+0.2°C
047.1°C
To proceed to the alignment of the sensors T1 and T2, see pg. 16.