Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast
Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc
Page 29 of 42
12- dPr
Too much moisture in the chamber before the sterilizing cycle starts: after having
opened the door, deactivate the alarm by pressing the selected program for 3 sec., then
repeat the cycle.
44- LLn
Contact the After-sales Service Center if the fault occurs again.
Check all the vacuum pump pipes. Replace them with genuine spares if they are
damaged.
Check all the solenoid valves to make sure that there are no leaks and clean and/or
replace the cores if necessary. NOTE: this operation must only be carried out by the
After-sales Service staff.
Check whether the pressure gauge pointer drops below –0.45 Bar because this means
that there is certainly a vacuum in the chamber and that the pressure switch setting
must therefore be checked.
19- StP
Connect the printer used to check and file the data.
20- H2o
Fill the tank with distilled or demineralized water using the supplied accessories.
39- Err
( For technicians only ): Repeat the microprocessor re-generating process.
42- drY
Supplementary drying cycle repeatable several times. To use when normal drying has
not been sufficient. Press DEDUCT with the door open to select the supplementary
drying cycle. The cycle lasts 9 minutes.
46- USd
Make sure that the can that collects used water is not too full.
If connections between the autoclave and drain pipe are incorrect.
Faults involving the drain pipe (clogged or partially blocked).
Remove the pipe from the direct drain and place it in the supplied can.
1 1 .
M A I N T E N A N C E
Carry out the maintenance operations with the machine at a standstill, after having
removed the plug from the power socket and after having allowed the components to
cool down.
The declaration of conformity to the standards in force is not valid unless the
maintenance operations described below are carried out.
11.1 General maintenance
The machine needs to be serviced in a particular way. Users are therefore advised to comply with the
below listed recommendations:
Periodically check the condition of the electrical system with particular attention to the connection
cable.
Contact our Technical Assistance Service or specialized personnel if faults are
discovered.
Periodically clean the door, the seal and take particular care when cleaning the
internal part of the
boiler
, removing any scaling with the damp sponge (supplied with the machine). Correct cleaning
of these parts of the machine will ensure that the sterilizing cycle takes place in the right way.
If the machine is not to be used for a long period of time, leave the door ajar and store it at a room
temperature of >8°C.
Use a damp cloth to clean the casing of the machine. Never ever use inflammable liquids for any
reason whatsoever.
The bacteriological filter in mod.
STERILIX VACUUM PLUS, STAR and STAR FAST
must be
changed at least once a year. New ones can be ordered with code 1067.