Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast
Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc
Page 32 of 42
1 2 .
F A U L T S
W I T H O U T
M E S S A G E S
1)
The machine fails to come on when the main switch is turned to position 1.
Remedy:
A) Check the power socket. If it fails to comply, have it replaced with a suitable one (16 A. 230 Volts).
B) Use a tester to make sure that the fuses installed behind the inside machine are functional. Replace
them if necessary, remembering to fully tighten the fuse-holder screws in the correct way.
C) Use a tester to check the voltage rating of the socket. We recommend the SCHUKO type (for
technicians only).
-------------------------------------------------------------------------------------------------------------------------
2)
Switch N° 10 comes on but there is no signal on the display.
Remedy:
A) Demount the casing and make sure that the fuses on the electronic board are well connected.
B) If the fuses on the board have burnt out they must be replaced with others of an equal value as shown
on the labels affixed near them.
--------------------------------------------------------------------------------------------------------------------------
3)
The cycle fails to start at once. High temperature in the chamber (between 90° and 130°) with
flashing sequence.
Remedy:
A) The cycle will automatically start when the temperature in the chamber has dropped to below 90°.