Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast
Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc
Page 6 of 42
Description
Unit of measurement
Values
Power source voltage rating
V
230 V.a.c. ~
Power source frequency
Hz
50
Machine power rating
W
1500
2500 (for Mod.
STERILIX STAR FAST
Boiler volume
mm
Ø 240 x 340
Operating pressure
Bar
2.1 - 1.1
Machine dimensions X, Y, Z
mm
510 x 450 x 390
Gross weight
Kg
49.5
Pack dimensions
mm
650x550x560
Operating temperatures
°C
+ 3°C… 35°C
Distance from walls
cm
8 – 10
X = depth
Y = width
Z = height
The electrical system is protected against the risk of over-currents installed inside
the machine by two 15 A delayed action fuses ( T 15 A ) for mod. STERILIX STAR
FAST and 10A ( T 10 A) for other models.
The electronic control and powering boards protect the various users by means of 3
dedicated fuses: 2 quick-acting 1 A types (F 1 A) and one 100 mA delayed action
type (T 100 mA) – (see enclosed wiring diagram).
Only have these fuses changed by a skilled electrician.
The machines have been designed to operate up to an altitude of 1200 mm above sea
level. Contact the manufacturer if the machine must operate at a higher altitude as
special models and/or devices may be required.
The noise level of the machine has been tested. The continuous equivalent weighted
acoustic pressure level A is less than 70 dB (A) and the continuous equivalent
weighted acoustic power level A does not exceed 85 dB (A). The measured noise
level measured is therefore very low. For this reason, the operator need not take any
particular measures (also in view of the fact that the machine operates
automatically).
DATA AND SPECIFICATIONS OF THE PRESSURIZED VESSEL:
Serial N° of vessel, Serial N° of door
Month N°, Year N° (see internal label)
Max. operating pressure:
2.2 Bar
Max. operating temperature:
160°C
Min. operating pressure:
-0.9 Bar
Test pressure:
8 Bar
Reference standards:
TRD 421
Volume of vessel:
15.5 l.