Radio Frequency Systems
BD4200W Series Air Dryer User’s Guide
Page 69 of 114
100518835 - Rev. B
9.5 Measuring Incoming Voltage
9.5.1
Remove the Top Cover (section 8.11 ).
9.5.2
Locate the Power
IEC
Connector inside the Top
Section of the air dryer.
9.5.3
Use a Voltmeter to measure
the voltage:
9.5.3.1
Place the probes
between the IEC
Connector and terminal
insulation so that they
touch the metal
contacts.
The voltage should measure:
110 - 125 VAC
for the BD4200W & BD4200WLP models.
220 - 230 VAC
for the BD4202W & BD4202WLP models.
If the incoming voltage measures less than indicated above, it is recommended
that steps be taken at your facility to increase the incoming power to the
recommended levels.
9.5.4
Reinstall the Top Cover (section 8.11 ).
WARNING!
Extreme care should be exercised to avoid contact with live
electrical circuits.
It is highly recommended that you remove all
jewelry before performing any procedures.