Testing and Care
Page 20 of 28
Baby Check Transmitters User Guide
0510-1124-I
2. Allow the cleaner to remain on the
transmitter for the amount of time
specified by the manufacturer. When
using surface disinfectants and
decontaminant cleaners, always follow
the manufacturer’s instructions.
3. Wipe the transmitter with a clean cloth
or towel, and then allow the transmitter
to air-dry. The transmitter can also be
rinsed with water, if required to in the
manufacturer’s instructions.
4. Place the clean transmitter in a storage
container for future use.
Testing
Do not use the transmitter tester on a
transmitter or clearing device with an
expiration date that has expired as the
results may not be accurate (the printed
expiration date on the device takes
precedence over the transmitter tester
results).
CAUTION
: You must test all transmitters
and clearing devices prior to use to verify
proper operation. Failure to test the
transmitters before use can result in system
failure, mismatch, and/or an abduction. In
addition, failure to test transmitters voids the
RF Technologies Product Warranty.
WARNING
: The Baby Check transmitter
CANNOT
be tested with the transmitter
tester used with the Infant transmitters or
Transmitter Clearing Device.
NOTE:
In order to avoid
nuisance alarms, cleaning
procedures must be
performed at least 100 feet
from any Alarming Band
Zone.