background image

A Mother Transmitter is a small, yellow device placed on the wrist of a new mother.

When it is activated, it sends a signal that is received by Exit Alarm Receivers and Alarming
Band Receivers. If a mother is handed a child wearing a transmitter that does not match her
transmitter, the Red LED on the Mother Transmitter flashes and an alarm sounds to signify a
mismatch.

Test all transmitters prior to use to verify proper operation.

To fasten a Mother Transmitter

1

Cut off 6-7" of alarming band material.

2

Open the clamps of the Mother Transmitter by gently pulling them outward,
one clamp at a time, from the top of the Transmitter.

3

Insert the alarming band material through the slot on one side of the Mother
Transmitter, from bottom to top, with the printed fabric side facing out. Guide
the alarming band material through the slot at the top of the clamp.

4

Snap the clamp into place.

11

Cleaning and Caring for a Mother Transmitter

2

Pre-clean the Mother Transmitter using an approved cleaner.

Use the provided brush to clean the hard-to-reach areas.

Brushes should be sanitized after each use according to the manufacturer’s
instructions and your facility’s infection control procedures.

3

Sanitize the Mother Transmitter by spraying it with an 

approved 

cleaner.

Using an approved cleaner does not confirm the suitability of the cleaner to
meet your infection control standard; it only indicates that the Mother
Transmitter will not be harmed by the cleaner.

Allow the cleaner to remain on the Mother Transmitter for the amount of time
specified by the manufacturer. When using surface disinfectants and
decontaminant cleaners, always follow the manufacturer’s instructions.

In order to avoid nuisance alarms, cleaning procedures should be

performed at least 100 feet from any Alarming Band Zone.

Use of non-approved cleaners on  Mother Transmitters may cause

permanent damage to the Transmitter and the alarming band piercing
contacts and will void any and all warranties.

This chapter provides detailed information about cleaning and caring for a Mother
Transmitter.

To clean a Mother Transmitter after use

1

Remove the alarming band from the Transmitter and dispose of the alarming
band material.

Approved 

cleaners:

Airex 109A
Cavicide™
Health-Tec™
Hi-Tor  Plus™
Isopropyl Alcohol
Neutral Quat 128™
Quik Fill 920™
Quest 256
Virex II-256
Sani-cloth HB (wipes)

Note: 

Some agents are intended for topical,

antimicrobial use only and will not destroy
viruses such as Hepatitis B and HIV.

Use care around the piercing contacts on the
Mother Transmitter.  Placing your fingers
directly on the piercing contact may result in
injury.

Bleach
Cleaners containing bleach
Cidex

™ (Including Activated, Plus, and OPA)

3M Phenolic

Virex TB
Virex HB

Non-approved

 cleaners:

DO NOT USE

6

Fastening a Mother Transmitter

Transmitter on a Mother

Summary of Contents for Safe Place Mother Transmitter

Page 1: ...User Guide P N 0510 1020 A Released 12 30 03 Mother Baby Match Mother Transmitter User Guide ...

Page 2: ... contacts on the MotherTransmitter Placing your fingers directly on the piercing contact may result in injury The transmitters may be sprayed or rinsed during cleaning but are not intended to soak for an extended period of time Other documentation to consult CA9450 Safe Place Mother Baby Software User Guide 0510 1014 CA9450 System Guide 0510 1018 Contents Preface 5 The MotherTransmitter 5 Using th...

Page 3: ...ANY OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING ANY IMPLIEDWARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE Part 0510 0121 C KIND WHETHER OR NOT FORESEEABLE OR IN CONTEMPLATION OFTHE PARTIES EVEN IF SELLER HAS BEEN ADVISED OFTHE POSSIBILITY OF SUCH DAMAGES ...

Page 4: ...y at its sole option instead elect to refund the purchase price paid for the defective product The criteria for all testing shall be Seller s applicable product specifications utilizing factory specified calibration and test procedures and instruments No allowance shall be made for local repair bills or expenses without the prior written approval of Seller If Seller elects to replace any part such...

Page 5: ...hild or resident sounded the alarm and the exit where the alarm occurred The system can automatically lock doors deactivate elevators and activate remote cameras The CA9450 System provides further security by alerting staff if a transmitter band is tampered with or if any wire in the system is cut or damaged About this Guide The MotherTransmitter User Guide is intended for users who fasten clean a...

Page 6: ...evice will beep once when you press the button If the Mother Transmitter is working properly the indicator light flashes and a tone sounds once every 15 seconds Wait for the flash of the indicator light and the tone from the Transmitter Tester to verify that the Mother Transmitter is functioning 3 If the Mother Transmitter fails this test check the battery in the device The Transmitter Tester requ...

Page 7: ...Central Server and Clients also identify which child sounded the alarm and the exit where the alarm occurred The Safe Place System provides further security by alerting staff if an InfantTransmitter s special alarming band is tampered with or cut The Mother Transmitter This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not caus...

Page 8: ...irm the suitability of the cleaner to meet your infection control standard it only indicates that the Mother Transmitter will not be harmed by the cleaner Allow the cleaner to remain on the MotherTransmitter for the amount of time specified by the manufacturer When using surface disinfectants and decontaminant cleaners always follow the manufacturer s instructions In order to avoid nuisance alarms...

Page 9: ...her Transmitter and inserting new clamps To replace a MotherTransmitter clamp Transmitter with both Clamps snapped closed Left Oval Shaped pin Right Hole Right Oval Shaped pin Left Hole 5 Place theTransmitter and alarming band onto the wrist of the mother 6 Insert the alarming band material through the slot on the other side of the MotherTransmitter from bottom to top 7 Again guide the alarming ba...

Page 10: ...tter A Red LED appears indicating that the MotherTransmitter is clear 2 After the transmitter beeps and lights single tap theTransmitter again This enables the tag so that it is ready to link the InfantTransmitters 3 Bring the infant with a fastened Infant Transmitter within 10 to 12 inches of the Mother Transmitter to link the two e d o M s p a T n o i t p i r c s e D e l d I n a o t y l e t a i ...

Reviews: