4.
RUN mode
Open the door by turning the door handle upwards and pull the door open.
Load the machine
Bedpans
are placed in the stationary turning stand in the door as show in fig. 1 when the stand
is in upper position. Bedpans are emptied automatically when the door is being closed.
Bedpan bucket
is placed on the grating turned down from the back wall of the wash chamber;
the mouth of the bucket must be to the left (2). Before placing the bucket on the grating it is
emptied into the machine. Lid and seating are placed between the supporting rails of the side
and back walls.
Urine bottles
(4 pcs) are pushed over the spray tubes in the door "stomach side" turned
upwards (fig. 1, 2, 3 and 4). Urine bottles can be washed together with bedpans, bedpan buckets
or wash pans. Urine bottles can not be washed when an instrument basket is being used.
Wash pans
(4 pcs) are placed into the gaps of the grating turned down (3).
Other objects
as kidney bowls are washed on the grating; they must
be placed upside down on the grating (4).
Suction bottles
are washed
on the special racks (fig.4).
Small
items are washed
in the basket (fig. 5).
Note !
Do not pour any liquid or other substances
into the chamber which may cause foaming.
Hint !
Foaming problems (Faults 11 & 10)
can be reduced by adding an extra
"cold water flush"-phase.
Close the door and lock the door by turning the door handle downwards
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
16