Operation

© Rice Lake Weighing Systems

 

 All Rights Reserved

9

3.4 Troubleshooting

Refer to the following table to check and correct any failure before contacting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead batteries

Replace batteries or connect to AC power

Faulty electrical outlet

Use a different electrical outlet

Bad power supply

Replace adapter

Questionable weight or the scale does not 

zero

External object is interfering with the scale Remove the interfering object from the scale
Display did not show 0.0 before weighing Help the patient off the scale, zero the scale and begin the 

weighing process again

Scale is not placed on a level floor

Ensure scale is level and begin the weighing process again

Scale is out of calibration

Check the weight with a certified calibration weight 

The display shows a 

STOP

 message

The load on the scale exceeds the 

capacity of the scale

Remove the excess weight and use the scale according to 

manufacture specifications

The display shows 

LO Bat

 message

The battery is low

Replace batteries

The display shows E and Err messages as detailed below
E06

Identifier - ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale

E4L

BAT

Battery low, but still usable- one bar left on indicator display

E4U

Battery low and unstable - no bars left on indicator display

E11

CAL

Calibration Error - recalibrate scale

Err 1

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly; Check the cables 

and mechanical connections; if the problem persists, replace 

the set of load cells

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight; Cannot calibrate

Check the load cell mechanical surroundings and ensure 

nothing is contacting the load cell and that the cables are 

properly welded

Err 7

Scale isn’t moving

Make sure feet are installed on the scale. Turn the feet all 

the way in and then back them out three full turns, then level 

the scale

SAT

Damaged load cell cable

Replace load cell cable

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Table 3-2.   Troubleshooting 

Summary of Contents for 150-10-6

Page 1: ...150 10 6 Digital Physician Scale Waist Level Software Version 11525 Technical Manual PN 162342 Rev B August 18 2022...

Page 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Page 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Page 4: ...Connections 6 2 5 Leveling the Scale 6 3 0 Operation 7 3 1 Weighing 8 3 2 Hold Release Function 8 3 3 Using the Body Mass Index BMI Function 8 3 4 Troubleshooting 9 4 0 Configuration 10 4 1 General N...

Page 5: ...is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful i...

Page 6: ...t the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any o...

Page 7: ...back Damage caused by improper packaging is not covered by the warranty 2 2 Scale Assembly Use the following steps to set up the scale A Phillips head screwdriver will be needed for installation 1 Pla...

Page 8: ...attach the column to the scale base as shown in Figure 2 3 Figure 2 3 Attach With Screws 2 2 1 Wheel Assembly Attach the wheel assembly to the scale prior to connecting the second column To attach th...

Page 9: ...lumn 2 Once the cable is connected place the upper column onto the lower column aligning the screw holes Ensure the cable does not get pinched Figure 2 6 Insert Upper Column into Lower Column 3 Secure...

Page 10: ...gure 2 8 Rice Lake Weighing Systems offers optional AC adapters utilizing an adapter not supplied by Rice Lake Weighing Systems voids all warranties Figure 2 8 Power Connection 2 5 Leveling the Scale...

Page 11: ...ff the scale A second press releases the weight value Not active while in BMI mode BMI Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient...

Page 12: ...to zero Pressing prior to the patient getting on the scale will also hold the weight display 3 3 Using the Body Mass Index BMI Function Use the following steps in determining the BMI LB Mode 1 Ensure...

Page 13: ...le E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on indicator display E11 CAL Calibration Error recalibrate scale Err 1 Load cell cab...

Page 14: ...Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 1 2 Enter Numbers Press to enter parameter va...

Page 15: ...cated in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pressed...

Page 16: ...6 Use the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be chan...

Page 17: ...umeric value and ARW If you do not want to change this value press the BMI key to move to the next setting Example from ARW to SAL If you want to change the value use the following steps 1 Press ENTER...

Page 18: ...een a numeric value and ATOUT If you do not want to change this value press the BMI key to move to the next setting If you want to change the value use the following steps 1 Press ENTER key to change...

Page 19: ...ant to change this value press the BMI key to move to the next settings If you want to change the value use the following steps 1 Press ENTER to change parameters 2 Use the Up Down arrow keys to selec...

Page 20: ...e back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover...

Page 21: ...le until ID flashes 2 Press Press to access the test menu 3 Continue to press to scroll through the various menu items 4 Once complete press until Done displays 5 Press to start the weighing process F...

Page 22: ...as the weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Table 5 1 Print Format T...

Page 23: ...rotocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC W02...

Page 24: ...D Z Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protoco...

Page 25: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a P...

Page 26: ...personnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for...

Page 27: ...table baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 F to 104 F 14 C to 40 C Storage Temperature 32 F to 158 F 0 C to 70 C Humidity 85 relativ...

Page 28: ...150 10 6 Digital Physician Scale 24 Visit our website www RiceLake com...

Page 29: ......

Page 30: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...