Appendix
©
Rice Lake Weighing Systems
●
All Rights Reserved
21
Shorting the excitation voltage shuts the excitation voltage off. The only way
to restore excitation voltage is to cycle power.
LINOFF
Low offset
Zero load at power up is less than initial zero range
(INIZR) setting of calibration zero
–
add the
missing load.
NOBATT
No battery
The RTC lost time/date tracking at previous power
off state due to low battery or no battery condition.
The printer, accumulator and AUDUT functions will
fail to get time and date.
EUCKSM
Configuration checksum
The checksum value of configuration has changed
from that stored in memory.
OIMLER
OIML parameter error
Parameter set incorrectly for use in the OIML
mode. Example: Primary units set for lb or oz.
EE-ACC
Accumulator error
Error with the accumulator such as attempting to
display an accumulated value greater than six
digits.
Error
Message
Description
Solution
Table 3-1. 480 Error Messages