DC-400 Counting Scale
6
Visit our website
2.4 Set Country
The country is set on initial use of the scale. Enter
Z mode
before setting country. Span Switch must be enabled to set country.
2.5 Calibration
Span Switch must be enabled. Enter
Z mode
before calibrating.
2.5.1 Calibration for Scale 1 (Internal Scale)
Operation
Function
Press and hold
and press
,
,
,
To enter Country Setting mode
Press
or
Select country
Press
Update country setting
Press
or
to skip this step
Load default SPEC setting
Press
Change scale type
Press
Update scale setting
Press
Select capacity
Press
Update capacity
Press
or
to skip this step
Clear Memory
Table 2-2. Set Country
Operation
Function
Press and hold
while pressing
,
,
,
Enter Calibration
Press
Ready to calibrate zero point
Confirm no weight is on scale, press
Calibrating zero point
Put weight on scale, press
Calibrating Span Point 1
Take weight off scale
Scale is ready for second weight
Place a second weight on scale, press
Calibrating Span Point 2
Take second weight off scale
Sale is ready for third weight
Place a third weight on scale, press
Calibrating Span Point 3
Take third weight off scale
Scale is ready for fourth weight
Place fourth weight on scale, press
Calibrating Span Point 4
Take weight off scale
Return to
Z mode
Table 2-3. Calibration for Scale 1