Setup
© Rice Lake Weighing Systems
●
All Rights Reserved
7
2.5.2 Calibration for Scale 2
Operation
Function
Press and hold
while pressing
,
,
,
Enter Calibration
Press
Select Scale 2
Press
Set weight decimal point
Press
or
Increase or decrease the number of decimal points
Press
Store weight decimal point setting, ready to enter platform capacity
Enter capacity, press
Store capacity setting, ready to enter minimum display
Enter minimum display, press
Store minimum display setting, ready to select Single- or Multi-Interval
Press
or
to select Single- or Multi-Interval and then press
Store interval setting
Enter Max1 setting, press
Store Max1 setting, ready to enter minimum display in weighing range Max1
The operation scenario only occurs on multi-interval
Enter Min D1, press
Store Min D1, ready to calibrate zero point
Reference weight for calibration must exceed or equal 10% full scale capacity
Confirm no weight is on scale, press
Calibrating zero point
Enter Span calibration weight, press
Calibrating span
Press
Return to
Z mode
Table 2-4. Calibration for Scale 2