Calibration
©
Rice Lake Weighing Systems
●
All Rights Reserved
11
3.2 Calibration
See the indicator Technical Manual to determine indicator calibration procedures.
1. Load the scale to near capacity two or three times to ensure everything is seated.
2. Ensure there is no load on the scale and place the indicator in the calibration mode.
3. Perform a zero calibration.
4. Place test weights equal to 70% – 80% of the scale capacity on the platform. When using several weights, distribute
them evenly around the platform.
5. Perform a span calibration.
6. Remove the test weights and check the zero reading.
7. Repeat the calibration process, if necessary.