8. Insert the plastic wire housing back
into the underside of the LED assembly
making sure it is fully inserted and
secure.
9. Replace the black plastic cover and
secure using the 4 foot screws removed
in step 2.
New LED bulb
Wire housing
Replacement 2W LED Assembly
Part #HS3-001
10. Return the microscope to its upright
position. Plug the microscope back
in, power on, and adjust the rheostat
control as needed.
LED Light Replacement